General Purpose Incubator BGI1O1 (BIGP-201)

Heating Incubator
  • Sea, Air, Door to Door Shipping
  • 1 Year Warranty
  • US & European Standards



General Purpose Incubator 220V 50Hz RT+5~65°C

Specifications

ModelBGI1O1
Electrical Requirement220V 50Hz
Temperature RangeRT+5~65°C
Display Resolution0.1°C
Ambient Temperature+5~35°C
Temperature Uniformity±1.5°C (at 37°C)
Power Consumption200W
Chamber Volume16L
Interior Dimension (WxDxH) mm250x260x250
Exterior Dimension (WxDxH) mm530x480x420
Shelves (pcs)2
Timing Range0-5999min
Remark

Description

Microprocessor Controller (with timing function)
Provided as a necessary equipment for scientific research to colleges as well as biological, agricultural and scientific research departments for storage of mould and biology cultivation.

Features

• With inner glass door for easy observation.
• Polished stainless-steel chamber.
• Independent temperature-limiting alarm system ensures experiments run safely.(Option)
• Printer connector and RS485 connector are options which can connect printer and computer to record the parameters and the variations of temperature.(Option)

Options:
• Printer
• Independent temperature-limiting Alarm system
• RS485 connector
• UV Sterilizer

Operating Manual

Download


1. Safety warning
2. Product introduction
3. Operation of the product
    1. Preparation before operation
    2. Power on.
    3. Temperature setting
    4. Timing setting
    5. Check temperature control accuracy
    6. High/low speed settings
    7. Methods for improving the accuracy of temperature control.
    8. Our suggestion
    9. Call out sequence of each function
    10. Use of optional accessories "independent temperature limit controller"
4. Appendix
    1. Technical indicators
    2. Table of Functional Parameters
    3. Trouble removal

Tags:, Description:, Attributes-End, Tips for ensuring safety, The matters contained here are extremely important and must Tips for ensuring safety

The matters contained here are extremely important and must be strictly observed.

1. Safety warning

Danger! The improper use of this unit may cause property damage and/or personnel injury.


1. This product must be reliably grounded and away from electromagnetic interference sources (The Zero line or the Neutral line should not be as the grounded line).

2. Please ensure the voltage and frequency of the power supply are compatible with the powerrequirements before use.

3. This unit must use the included electrical cord with a dedicated electrical circuit with a confirmed electrical ground connection.

4. The power switch must be in the "OFF" position when power is connected or disconnected from unit

5. Do not arbitrarily lengthen or shorten the power supply connection wire.

6. Unauthorized repair is not allowed and such authorized repair should be carried out by the special personage.


! Warning ( it may cause property damage or personal injury)

1. Make sure to read and understand thoroughly the Product's Operating Instructions before the operation may be carried out.

2. To take out the power plug, make sure not to pull directly the power cable.

3. 304 stainless steel liner is not acid resistant, please pay attention to anti-corrosion measures. Do not use acidic media inside the box!

4. The power cord must be removed from receptacle when any of the following occur:

4.1 When replacing the fuse;

4.2 When the product is waiting for overhaul due to faults;

4.3 When the product goes out of service for a long time;

4.4 When the product is being moved.


! Caution (it may influence the lifetime cause the abnormal operation of the product)


1. The product should be located on the solid and hard surface to keep it in a horizontal mode.

2. Keep certain space around the product

3. The product must be used in the specified conditions. (The detail refer to chapter 3, item 1.1~1.4).

4. The product must be used in the correct setting mode. (The details see chapter 3)

5. Do not open/close the product box door by gravity, otherwise it may easily cause the box door to fall off, damage the product, and cause injury accidents.

2. Product introduction

1. Outline drawing (This drawing is for reference only, subject to the actual object!)

1. The product should be located on the solid and hard surface to keep it in a horizontal mode., 2. Keep certain space

Figure 1: Desktop computer (applicable to BIGP-201model 1. The product should be located on the solid and hard surface to keep it in a horizontal mode., 2. Keep certain space

Figure 2: Desktop computer (applicable to BIGP-203-BIGP-206 models)


Figure 3: Vertical machine (applicable to BIGP-501-BIGP-503 models), 1. Controller, 2. Box door, 3. Observation window, 4.

Figure 3: Vertical machine (applicable to BIGP-501-BIGP-503 models)

1. Controller

2. Box door

3. Observation window

4. Door handle

5. Casters

6. Independent temp limiting controller (optional

7. Power switch

8. Damper knob

9. Fuse holder

10. Power cord


2. Overview of structure and function

DHP series electric thermostatic incubator consists of:

Box - The outer shell is made of steel plate, and the surface is sprayed with plastic; The inner liner is made of mirror surface rusty steel, and the four corners of the semicircular arc are easy to clean. The shelf spacing inside the box is adjustable.

Heating - Electric heating.


Circulation - The box is equipped with a motor, which operates through a fan to force the smooth circulation of gas in the working chamber to improve temperature uniformity. At the same time, it has the function of fast and slow speed change.


Control - Microcomputer type electronic controller, sensor using Pt100 platinum resistance temperature control accurate, stable and reliable.


Protection - With multi-level independent overtemperature protection and alarm system:

One level overtemperature protection: The controller is equipped with an upper deviation alarm function, which allows users to adjust the deviation AL1 parameter as needed and set it by themselves.


Second level overtemperature protection: When the actual temperature of the equipment exceeds the set protection value for the high temperature limit, the power supply to the heating section is automatically cut off, and an audible and visual alarm is issued.". Due to the fact that the circulating fan still operates normally during an alarm, the alarm will automatically be cleared when the temperature is lower than the set value. (User option)


Convenience -(BIGP-201-BIGP-202) Direct observation through the inspection window of the box door, convenient and clear. While (BIGP-203~BIGP-206) have dual interior and exterior box doors, which can be opened for observation through a glass interior door. When the doors of the desktopBIGP-203~BIGP-206 box and the glass doors of the BIGP-203 - BIGP-206 box are opened, the breeze circulation automatically stops, reducing the temperature "overshoot" caused by opening/closing the box door. The vertical BIGP-501~BIGP-503 doorless switch does not have this function.


(BIGP-201-BIGP-206) models can be optionally equipped with a liquid crystal display controller or a multi stage programmable liquid crystal temperature controller. (BIGP-201-BIGP-206) Type B is a liquid crystal display controller (see VI, Instructions for the Use of Liquid Crystal Display Controllers), optionally equipped with a multi stage programmable liquid crystal temperature controller.


Option Name

Introduction to optional features

Picture

Intelligent programmable liquid crystal temperature controller

▪ Microcomputer program controller, time and heating rate.

▪ Can preset 7 groups of 9 steps programmable, with a set time of 1 to 99 hours and 59 minutes per segment.

▪ Multi stage programmable control can simplify complex test processes and truly achieve automatic control and operation.

BIGP-201

RS485/232 Communication interface and software CD

Display and monitor various experimental data by connecting a computer.

BIGP-201

Embedded printer

Print the data required for the experiment.

BIGP-201

USB flash disk storage data

USB data transfer interface, which can be used for data transfer.

BIGP-201

Add BOD socket

Special socket for BOD instrument

Independent temperature limiting controller

Automatically interrupt when the temperature exceeds the limit, ensuring safe operation of the experiment without accidents

BIGP-201 or BIGP-201

Ultraviolet sterilization system

Wireless alarm system

Wireless alarm system (SMS alarm system)

Test hole

25mm/50mm/100mm Test hole

It can be installed on the side of the box and can be wired according to the needs of the placement site.

Note: Choose one from RS485/232, printer, and USB flash drive.

3. Operation of the product

1. Preparation before operation

1.1 The products should be used work normally under these following conditions:

a. Ambient temperature: 5°C~35°C; Relative humidity≤85%

b. Atmospheric pressure:86106KPa

c. Elevation not higher than 2000m

d. Power supply:220±22V 50±1Hz

e. Keep away from heating sources, no intense shock source and strong electromagnetic field around.


1. PV display unit (Red), * Display measured value., * Dispalys the various characters depending the statue of the

2. Power on.

1. PV display unit (Red), * Display measured value., * Dispalys the various characters depending the statue of the

1. PV display unit (Red)

* Display measured value.

* Dispalys the various characters depending the statue of the instrument.

2. (SV) display unit (Green)

* Displays the set value * Dispalys the various parameter values depending the statue of the instrument. Indicators

* COLD(AL2)Output light (yellow), lit when motor is operating at high speed.

* ALl Deviation alarm output light (red), lit during operation.

* AT The lamp (yellow) is shared with the TIME lamp, when lamp flashes, the meter runs at a fixed time; The table performs self tuning when it is normally on.

* HEAT heating output lamp (green) It shines when working for output.

4. SET Function key

* Setting value modification

* Call out of parameter symbols and confirmation of parameter modification.

5. Shift key

6,7. Number plus and minus keys

* "↑"Used to adjust parameter values or observe timing remaining time/timing recovery.

* "↓"Used to adjust parameter values or enable/disable self setting status/current real-time power, i.e. SV displays "o 100".


2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured temperature, and the SV screen displays the set temperature.

2.2 Close the glass door and the box door, and the motor runs. At this time, the incubator enters the working state.

3. Temperature setting

2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured

Figure 5

3.1 See Figure 5Click"SET"key, The PV screen displays " 2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured ", SV screen displays the set temperature value.

3.2 Use "←" to cycle through the flashing bits displayed on the SV screen.

3.3 Use "↓" or "↑" to change the value of the flashing bit on the SV screen until the desired temperature value is reached.

3.4 Click again "SET" key, Return to operating mode (PV screen display measure the temperature, SV displays the set temperature and enter the working state.

4. Timing setting

2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured

Figure 6

4.1 See figure 6, After pressing the SET key for 4 seconds, the PV screen displays " 2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured ", if SV screen display "0", it indicates that timing function not set (factory status).

4.2 Press the SET key several times, until PV screen display"L K ", trough 2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured key make the SV screen display 3 (timed unlock), press again SET key, PV screen return" 2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured "trough 2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured key, set the required value for timing (timing range: 1 to 9999 minutes), press the SET key again for 4 seconds, and the controller returns to the working mode. The timing starts to run. At this time, the AT light flashes, entering the calculation time. When the timing ends, PV and SV windows are displayed"End", the AT light goes out and the buzzer beeps 4 times to indicate a reminder.

4.3 Timing recovery: Press and hold the key until 2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured the instrument enters the working mode, and still operate at the original set timing time.

4.4 Self setting is not allowed after a timed start. To observe the running time, press the key 2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured once, and the SV screen displays the remaining time value. (Note: The determination of the total timing time should consider the combined calculation of temperature rise and constant temperature in two stages).

4.5 During timed operation of the product, it is allowed to modify the timing time "ST". The previous accumulated running time is "memorized" and runs to the new timing time. The product stops heating, and the buzzer beeps. (When the new timing time is less than the previous cumulative operating time, the heating output is immediately turned off and the buzzer sounds.)

4.6 When the power is cut off during operation or the power is turned off and then restarted, the timing needs to be reset.

5. Check temperature control accuracy

5.1 Select a digital thermometer with a resolution of 0.1°C that has passed the verification and is within the verification period as a standard and put it into the product work room, and ensure that the sensor temperature sensor is in the geometric center of the work room.


5.2 Choose a point within the temperature control range of the product and set the SV temperature control value. When the PV measurement value is equal to the set value, keep it at a constant temperature (1~2) hours or so (depending on the product specifications, the constant temperature time may vary) , Observe that the difference between the actual measured temperature value of the standard thermometer and the measured value displayed by the controller PV should be ≤±0.5°C.


6. High/low speed settings

6.1 Yellow lamp(COLD/AL2)light, high speed operation of fan.

6.2(See figure 7)When the temperature reaches: PV = SV +Fn,

That is, from high speed to low speed operation, and

the yellow light goes out at this time.

5.2 Choose a point within the temperature control range of the product and set the SV temperature control value. When the PV

Figure 7

7. Methods for improving the accuracy of temperature control.

7.1 Error correction between measured temperature and actual temperature in the working room.7.1.1 Press "SET" key more than 4 seconds (enter menu B), when PV screen display" 2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured "just release, press SET key Several times, find the "LCK" prompt.

7.1.2 Press " 2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured "key, make the SV screen display 18 (unlock).

7.1.3 Press again "SET" key Several times, find the "Pb" prompt for the control parameter that needs to be adjusted, and press the function key to display the control parameter as the desired value.


Pb = Original Pb value +Actual temperature - PV display value


7.1.4 Calculate the "Pb" value according to the above formula, and input it. (If the adjustment is not correct at one time, it can be repeated multiple times until it meets the standard.).

7.1.5 If Pb has been adjusted to the maximum (out of range) and cannot be adjusted again, set "Pb" to 0 and return to the working state.

7.1.6 Press the "SET" key, and "PK" will appear on the PV screen. Modify the "PK" to 1.000, and then return to the working state.

7.1.7 Select 1 test point within the temperature range, such as:PK=4000*Standard values of the recorder - displayed temperature measurements/Displayed temperature measurement (be sure to retain three decimal places)

7.1.8 Set a new "PK" parameter value, and if it is no longer accurate, adjust the "Pb" according to the steps in 7.1.3 of this chapter (repeatedly until it meets the standard).

7.2 Suppression of temperature instability (causing up and down fluctuations or excessive overshoot).

7.2.1 Refer to 7.1.1 of this chapter to enter the B menu.

7.2.2 Increase the parameter "P" (proportional band) by about 0.5 each time.

7.2.3 Simultaneously change the parameter "I" for about 100 seconds each time, and simultaneously change the parameter "D", D=I/4, and observe the temperature again until the temperature overshoot meets the requirements.

7.2.4 If it still cannot be eliminated, refer to Section 7.1.6 of this chapter to enter the B menu.

7.2.5 Find "HP" (output power) and reduce it by about 10% each time.

7.2.6 Self setting startup/shutdown: Hold the key 2.1 Turn on the power switch. At this time, the power indicator light is on, the controller PV screen displays the measured for a long time, the AT light will be on for a long time during self setting, and the SV window will alternately display the SV value and the words "- AT -"; At the end of self setting, the AT light goes out.

7.2.7 When the required temperature is low, a secondary setting method can be used to reduce temperature overshoot.

(For example, when the required temperature is 40 °C, set 38 °C first, and then set 40 °C after the temperature is basically stable, which can reduce or even eliminate temperature overshoot.)

8. Our suggestion

8.1 When the set temperature "SV" is close to RT (ambient temperature)+5 °C, use the secondary setting method (see Chapter 3, Article 7.2.7)


8.2 If the humidity in the box is found to be too low during work, a water pan can be placed inside the box. Through natural evaporation, the relative humidity can generally reach about 90% (the water pan only provides a humidity environment and cannot control humidity).

9. Call out sequence of each function

8.2 If the humidity in the box is found to be too low during work, a water pan can be placed inside the box. Through natural

10. Use of optional accessories "independent temperature limit controller"

The over-temperature protector is an independent protection system. When the temperature controller fails and the temperature is out of control, when the temperature in the working room reaches the temperature limit setting value of the over-temperature dial, the over-temperature protector will automatically cut off the heating and sound an alarm.

(As shown in the right picture) When the temperature in the working room is lower than the temperature limit.After the set value, the protection system is eliminated and the instrument resumes work.Repeat this way until the fault is eliminated.


BIGP-201


The specific operations are as follows:

6.1 The temperature limit setting value should be greater than or equal to • (SV+AL) + (10~15)°C

6.2 See Figure 2, use the + and-buttons on the over-temperature setting dial on the panel to set the required temperature limit.

Example: SV=130°C, AL=10 Then 150°C should be set up.

The specific operations are as follows:, 6.1 The temperature limit setting value should be greater than or equal to •

4. Appendix

1. Technical indicators

Model

BIGP-201 BIGP-7601

BIGP-202 BIGP-7602

BIGP-203 BIGP-7603

BIGP-204 BIGP-7604

BIGP-204 BIGP-7604

BIGP-204 BIGP-7604

BIGP-501

BIGP-501

BIGP-503

Power Voltage

AC 220-240 V 50-60HZ

Range of Controlling Temperature

RT (room temperature) +5°C -- 65°C

Precision of Controlling Temperature

±0.5°C

Output Power

200W

200W

300W

400W

600W

750W

1100

1400

2200

Volume

16L

35L

50 L

80L

160L

270L

420L

620L

1000

Working Size(mm)

250x260x250

340x320x326

415x360x355

500x400x400

500x500x650

600x600x750

640x585x1355

840x600x1355

1000x600x1600

Appearance Size(mm)

530x480x420

620x490x490

690x500x500

760x580x420

630x770x770

890x730x960

780x750x1880

980x800x1880

1140x800x2150

Note: "B" refers to the liquid crystal display controller.

2. Table of Functional Parameters

2.1 Trouble removal

LCK=3

Parameter name

Function

Setting range

ST

Timing

0~9999

The timing start method is optional. Timing is a countdown operation mode. Press the △ key to observe the current remaining time. Note: The timing value meter does not remember, and this setting is valid.

When the timer expires, the buzzer beeps continuously by pressing a short beep for four times. (The beep stops after three minutes.) The PV display window and the SV display window both display End. You can press and hold the plus key to reset the timer (with two beeps). After resetting, you need to reset the timer value. After a timed start, the TI timing light flashes.

LCK=18

AL

Upper and lower limit deviation alarm

0~100.0

PV>SV+AL cuts off the output and beeps an alarm, and the SV window displays E002.

PV<SV-AL beep alarm, SV window displays E006, and the output is not cut off. (First power-on exemption for lower limit deviation alarm)

The beep alarm can be muted by pressing any key.

Fn

High and low speed conversion value of fan

0~300.0

The PV is at a low speed within the range of SV ± Fn, and at a high speed outside the range. (with a return difference of ± 0.2 degrees).

Pb

Display value translation correction

-100.0~100.0

PV=PV+Pb

ATU

Self tuning

0 / 1

Note: The status of AL1 and AL2 will not change during self tuning.

It is recommended to turn on self-tuning when PV<<SV.

After opening the self-tuning, the SV window displays - AT -.

P

Proportional band

0~100.0%

Proportional action adjustment: the larger the P, the larger the proportional action range, and the lower the system gain. P=0 is a bit control.

I

Integral Time

1~4320 second

Integrated time constant. I larger, then the integrated action is smaller.

d

Derivative Time

1~4320 second

Derivative time constant. D larger, then the derivative action is also larger.

t

Heating period

1~60 second

The heating cycle is generally 2-3 seconds for the thyristor and 20-40 seconds for the relay output.

PK

Slope correction

-1000~1000

PK=4000*(Standard value of recorder - displayed temperature measurement value)/displayed temperature measurement value

HP

Output power suppression

0~100%

* Note that if the HP value is too small, the temperature will not reach the set value

LK

Key lock

0~255


2.2 B Menu.


LCK=28

Parameter name

Function

Setting range

Ad

postal address

0~128 0:broadcast

bA

Baud rate

0:1200 1:2400

2:4800 3:9600

4:19200

Fo

Communication format

3:8,N,2; 4:8,E,1

5: 8,O,1;

Pt

Print Interval

0-9999

yr

Year

21-50

MH

Month

1-12

dA

Day

1-31

Hr

Hour

0-23

Mi

Minute

0-59

on

Date modification confirmation

0-1 To modify the date and time of printing or USB flash disk storage, after modifying the above time, change this parameter to 1.

※Products are strictly tested before leaving the factory. When the technical indicators meet the requirements and work normally, no correction is generally required.

3. Trouble removal

S.N.

Failure

Reason for presuming

The handing ways for failure

1

No power when starting

The power outlet has no current or badly connecting

Checking and repairing

Coming line interrupting of power.

Changing it

Switch of the power doesn't open or run

Open (closing) the switch, changing it

Not installing the cartridge fuse or burn out

Installing the right cartridge fuse, and checking the reason, then starting after repairing

2

PV screen displaying….

The instrument or the sensor PST00 does not work

Changing it, PST00=0°C ,it is 100 Ω

3

The temperature not rising or after at some degree, not ring, but dropping

The setting temperature is lower the RT (surrounding temperature)

Resetting SV≥RT+5 °C

The timing of the instrument booting T≠0

Setting ST=0

The inner glass window is closed, but not closing the switch of the door

Adjusting the inner door's locking

The "heating light" temperature controller is on

The controlled silicon or the integrated piece does not work, changing it

There is voltage between two sides of the heating tube

The connector of the heating tube is hurt or dropped.

When over 60 °C, temperature suddenly not rising

The heat protection of the temperature relay, after nature cooling, it could restore.

4

Big difference of the displaying and actual temperature

Refer to the specification, and repair it.

5

Producing the static difference

Displaying temperature > setting temperature

Reducing the RT temperature, and modifying the "OUT-",

Displaying temperature < setting temperature

Increasing the "OUT-"

Blower fun does not run

Repairing it

The sensor (PST00) connects not well

Connecting it once more.

7

The noise of the motor is big

The vane of blower fan touching the wind way board

Placing the gasket to increasing the height.

The motor is blocked or no grease

Change it

6

Temperature could not be controlled

The heating light of temperature controller is on and not off

Changing the temperature controller

When heating light of temperature controller if off, but the temperature is still rising

Changing the dual controlled silicon BTA16


Packing list

S.N.

Class

Name

Unit

Amount

Remark

1

File

Operating specification

Set

1

2

File

Packing list

Set

1

3

Spare parts

Shelf

Piece

2

4

Spare parts

Core for the fusible cutout

piece

2

The things listed in this list are consistent with practicality. Packing operator: 2