Stability Test Chamber BGI1BG13 (BTST-1513)
Medicine Stability Testing Chamber- Sea, Air, Door to Door Shipping
- 1 Year Warranty
- US & European Standards

Specifications
| Model | BGI1BG13 |
| Temperature control range | 0~65°C without illumination15~50°C with illumination |
| Temperature Stability | ±0.5°C |
| Temperature Uniformity | ±2°C |
| Humidity Range | / |
| Humidity Stability | / |
| Illumination | 0~6000LX adjustable |
| Illumination difference | ≤±500LX |
| Timing Range | 1~99 hours each period |
| Humidity and temp adjusting | Balance temperature adjusting |
| Cooling system/cooling mode | Two sets of imported compressor work rotationally |
| Controller | Programmable (LCD screen) |
| Sensor | Temp: Pt100 |
| Ambient Temperature | RT+5~30°C |
| Electrical Requirement | AC220V 50Hz |
| Operating Power | App.600W |
| Rated Power | 3200W |
| Chamber Volume | 400L |
| Interior Dimension (WxDxH,mm) | 600x580x1140 |
| External Dimension (WxDxH,mm) | 860x855x1810 |
| Shelves | 3/4(4pcs) |
| Safety Device | Compressor overheating and overpressure protection, Fan overheating protectionOver temperature protection, Overload protection, Water protections |
| Remark | 1. Build in Embedded Printer.2. Paperless recorder (Option).3. GP series products have installed intensity of illumination detector. |
Features
Microprocessor controller (with timing function)• Fluorine-free design, R134a refrigerant.
• Microprocessor control, 304 stainless steel chamber, semicircular arcs at corners for easy cleaning
• Even air circulating system.
• 2 imported compressors and fan motor, 2 cooling systems ensure experiments run safely in long time.
• Over temperature and temperature difference alarms, humidity difference alarms.
• Independent audible and visible temperature-limiting alarm system ensures experiments run safely.
• Imported humidity sensor which can be used in high humidity environment.
• Balance temperature and humidity adjusting system.
• There is a 25mm instruction connection hole on the left side of the chamber for easy testing operation and temperature measurement.
• UV light system for periodic sterilization of chamber.(Option)
• RS485 connector can connect computer record and inspect the parameters and the variations of temperature.(Option)
Programmable Touch Screen
• Large LCD screen to display more data at same time
• English operation menu, display current data curves
• 100 groups with 1000 periods 999 circulations, max timing for each period is 99 hours 59 minutes.
• Auto lock after setting data.
Safety device:
• Compressor over-heat protection
• Fan over-heat protection
• Over-temperature alarm system
• Over-compressing protection
• Over-load protection
• Water lack protection
Storage conditions for long-term retention sample stability test
Temperature: +25°C ±2°C
Humidity: 60±5%RH
Time: 12 months
Storage conditions for accelerated stability test
Temperature: +40°C ±2°C
Humidity: 70±5%RH
Time: 6 months
Illumination under strong light conditions: 4500±500LX
Above related parameter is for reference only.
* With model "P" with Programmable Touch Screen
* With model "G" with Light