Muffle Box Furnace BFX1D5 (BFMF-1300-40)

Standard Muffle Furnaces
  • Sea, Air, Door to Door Shipping
  • 1 Year Warranty
  • US & European Standards



Muffle Box Furnace 1300 °C 40L Muffle Box Furnace 1300 °C 40L

Specifications

ModelBFX1D5
Working Temperature1300 °C
Volume40L
Inner Chamber Dimensions [mm] WxHxD300x300x445
Product Outer Dimensions [mm] WxHxD630x785x840
Net Weight [kg]112
Power [W]7500
Max. Current [A]3x11.4
Electrical Connection3 Phase
Heating ElementFe-Cr-Al
Thermocouple TypeS Type
Heating Element PlacementEmbedded into brick walls
Inner Insulation MaterialInsulating Fire Brick
Front Face Insulation MaterialInsulating Fire Brick
Door Insulation MaterialCeramic Fibre Board
Housing MaterialSteel Sheet
Housing CoatingEpoxy powder coating
ChimneyStandard
Heating Element ProtectionNone
Lockable Door HandleSidewards
Gross Dimensions [mm] WxHxD690x1000x845
Gross Weight [kg]136
SoftwareBASIC B2
Software based PID ControlYES
Display4 lines
Heating Program with4 steps
Custom Preset Program2
Heating Rate °C/min.1-20
Date & TimeYES
Auto Start at Certain DateNO
Show Remaining Waiting TimeNO
Skip the Waiting StepNO
Temperature Calibration via Menu (±10°C)NO
Sound Warnings at Step ChangesYES
Sound Warning at the End of the ProgramYES
Total Working Hour CounterYES
Calculator for Average Working TemperatureYES
Instantaneous Energy Consumption IndicatorNO
Over Heating Cut OffYES
Open Door SensorOPTIONAL
PC Connection KitOPTIONAL
Warranty Period2 Years
Note* Only 5th memory supports 32 levels **Independent for each level

Features

• Suitable for every kind of processes like ashing, sintering, preheating, melting, oxidizing.
• User friendly operating menu in 4 line display for easy operation.
• Durable inside lining made of light insulating bricks.
• Accurate measuring and controlling the temperature using K and S Type thermocouples.
• ±1 °C control accuracy.
• Designed for minimal space requirements in the laboratory, thanks to the highly engineered insulation design.
• Dual shell housing for low outer surface temperature.
• Optional protective gas connection, door switch, second thermocouple for different adjustable purposes.
• Lift up and Sidewards opening door possibilities.
• Standard chimney with fan.
• Robust, durable and reliable.
• Ideal for all washing, burning and similar processes.
• Bottom, side and top walls and front facing insulations made of light firebrick.
• Lockable door handle.
• Optional Lift-up door.

Operating Manual

Download


1. DEFINITION
2. SAFETY
    2.1 PROPER OPERATION
    2.2 OPERATION IN SAFE CONDITIONS
3. FIRST HEATING
4. PID CONTROL UNIT PROPERTIES
5. DISPLAY & MENU
6. MENU STRUCTURE
7. OPERATING BY MENU AND STARTING THE FURNACE
    7.1-ENTERING MENU
    7.2 TO ENTER & TO CHANGE VALUES
8. START AND STOP PROGRAM
APPENDIX-1 Technical Specifications
APPENDIX-2 Control Unit Features

1. DEFINITION

BFMF-1200-20 is a laboratory type heating furnace. Heating is performed up to 1300°C directly and/or by stepped control procedure according to selected model.

2. SAFETY

BFMF-1200-20 heating furnace may harm operating people or surrounding any other materials, unless this operating manual is applied during operating process.

Thus;

-Periodical maintenance should be applied.

-Cautions against accidents should be main concern of operators.

-All operating directives, warnings and recommendations in this operating manual have to be followed and applied carefully.

Unless operating directives, warnings and recommendations in this operating manual is followed and applied, company Biolab is not dedicated for the accidents that may occur.

2.1 PROPER OPERATION

BFMF-1200-20 can reach up to 1300°C according to selected model. Thus materials that can stand up to 1300°C can be heated in the chamber. The conditions that should be considered during operating or the points which are the reason no to start operational process are as follows,

• Furnace is not started, if there is any living thing in the chamber.

• Any burning, flammable, exploding, poisoning, (Benzene, LPG, Acetylene etc.) material that may harm when it is heated etc. is not stored or put inside the chamber. These types of materials should be kept away from the furnace.

• Outside case of the furnace may be hot. It should be considered. Especially, when the furnace is heated above 1000°C and if it is kept above this temperature 30 minutes or more, you shouldn't touch to outside case/surfaces without using gloves.

• Instant high heat should be considered, when the front door is open.

• Electronic and/or Electrical components may create induction current or magnetic field. It may harm any electronic equipment surrounding the furnace. Especially cardiac pacemaker users should be away from the furnace.

• It should not be operated in closed environments such as in cupboard etc.

• Furnace should not be operated by multi plug in tools. It should be operated by stationary plugs which are mounted on the wall.

Tags:, Description:, Attributes-End, BFMF-1200-20 is a laboratory type heating furnace. Heating is performed up to 1300°C - Explosive, flammable, burning, poisonous materials, don't heat up.

DANGER

Tags:, Description:, Attributes-End, BFMF-1200-20 is a laboratory type heating furnace. Heating is performed up to 1300°C - Hot Surface

WARNING

Tags:, Description:, Attributes-End, BFMF-1200-20 is a laboratory type heating furnace. Heating is performed up to 1300°C - Instant high temperature, when the front door is opened.

DANGER

2.2 OPERATION IN SAFE CONDITIONS

"Contribute in to SAFETY CAUTIONS in the work shop."

Please obey to safety rules. Please inform responsible person, as soon as you notice any abnormal condition for the furnace.

Tags:, Description:, Attributes-End, BFMF-1200-20 is a laboratory type heating furnace. Heating is performed up to 1300°C

Tags:, Description:, Attributes-End, BFMF-1200-20 is a laboratory type heating furnace. Heating is performed up to 1300°C

Please use protective glass or protective gloves, when you are working.

Don't let non-permitted person contact to any electrically alive parts. Possible electrically alive parts should be processed by responsible persons and by experts. Cables should be protected against/away from any heat, oil, oily material, sharp tools and materials. Cables should be kept away from furnace surface. Any cable may kill, if any parts of the cable are broken / damaged or cable touches on to the furnace surface. Any broken cable should be replaced by the new one.

Tags:, Description:, Attributes-End, BFMF-1200-20 is a laboratory type heating furnace. Heating is performed up to 1300°C - Danger of life due to electrical shock.

DANGER

Tags:, Description:, Attributes-End, BFMF-1200-20 is a laboratory type heating furnace. Heating is performed up to 1300°C - Danger of life due to electrical shock.

DANGER

To touch in to the electrically alive parts may kill, if cable is broken or damaged. Be aware of the environmental affects: Don't use electrical tools, equipment and machines in wet circumstances. Keep light intensity enough to make easier for the working people. Plug off the machine, before any cleaning, repair and maintenance process.

3. FIRST HEATING

Following conditions should be considered, directions must be applied step by step, and the importance of the first heating process should be kept in mind, when BFMF-1200-20 Furnace is operated and heated up for the first time.

3.1-Enough free field, surrounding the furnace should be kept. Recommended distance is minimum 30 cm. This free space will increase the furnace performance because of air circulation. At the same time, any fire or explosion danger which may occur due to any material left around the furnace. Please don't forget, when the furnace reaches to high temperatures, the surface of the furnace reaches to high temperatures too. Any flammable and explosive material which is close to furnace can be a reason for the fire or explosion danger.

3.2-Electricity plug should have proper ground connection and plug should have proper capacity (Cable thickness) according to maximum ampere of the furnace.

3.3-Some gas and fumes due to some chemicals which is used on the electronic parts, in the isolation materials and due to outer furnace surface paint can be harmful if it is breathed directly, during the first heating process. Protective glasses and breathing masks should be definitely used, during first heating and very good air circulation in the working room should be kept.

3.4-During the first heating up process, furnace should be heated up according to following temperature-time table

Heating up to 1100°C

Waiting for 1 hour

This procedure can be completed by BFMF-1200-20 Programs easily.

Please don't touch furnace without gloves, especially during first heating process, due to hot surfaces.

4. PID CONTROL UNIT PROPERTIES

Following properties are included in S16 model. Please check the differences between P - PX- B2 - E4 - U8 - S16 models at the page appendix 2.

• PID control

• Up to 32 stepped heating control

• 1 - 20°C/min heat acceleration control

• Up to 6 memory, easy working via saved process

• Sound warning during step changes & at the end of process

• Remaining time indicator which may be required during the step where temperature is kept at the same level

• Date and time

• Total working time indicator

• Auto start by setting date and time

• Indicator for the temperature of control PCB and warning for the overheating of control PCB

• User friendly control menu by only 6 buttons

• User friendly 4 line display

• First start-warranty time saving mode

• Thermocouple failure warning

• Automatic stop control over 1100-1300°C

• Door switch (Heating is cut off, when the door is open. Optional)

• 2 Years Warranty without Heating Elements, Thermocouple, Consumables and Freight Cost.

5. DISPLAY & MENU

Tags:, Description:, Attributes-End, BFMF-1200-20 is a laboratory type heating furnace. Heating is performed up to 1300°C


1-On / off button

2-Led indicating on going heating process

3-Led indicates that furnace works

4-Display

5-Heating start

6-Heating stop

7-Menu process input/enter/change

8-Menu process selection

9-Menu process selection

10-Menu options

6. MENU STRUCTURE

START-10 AUTO START

20 SYSTEM STATUS- 30 PROGRAMS

Heating Program -1-40 PROGRAMS

Heating Program -2-50 PROGRAMS

Heating Program -3-60 PROGRAMS

Heating Program -490 ERROR LIST

7. OPERATING BY MENU AND STARTING THE FURNACE

7.1-ENTERING MENU

After the furnace is started, if "MENU/BACK" button is pressed, it is possible to reach menu options. "▲" ve "▼" buttons are used to change menu options.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

MAIN MENU

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

DISPLAY FOR MENU OPTIONS

If you see "ENTER" at the right hand corner, it is possible to reach sub menu options via pressing "ENTER" and "▲" ve "▼" buttons are used to change options and "MENU/BACK" button is used come back to top menu.

7.2 TO ENTER & TO CHANGE VALUES

If you see "ENT" at the corner of present menu, value is entered & changed by "ENTER" button.

Changing value is between "> <" signs. "▲" ve "▼" buttons are used to increase and decrease the values. "ENTER" button enables the user to reach options which can be altered.

Altered values are saved, when "> <" signs disappear via pressing "ENTER"

You may come out the present menu and you can reach to top menu by "MENU/BACK" button.

STARTING

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

When START/STOP button is pressed, following display appears together with 'Bip' sound.

Menu below is traced to change furnace language as soon as the Furnace starts following display appear on the screen.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Changing furnace language is enabled by pressing "▲" ve "▼" and ''ENTER" buttons.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Then, main menu here below appears at the display his is main display to follow menu options and change the values. "MENU/BACK" button enables to enter menu.

10 AUTOMATIC START

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

AUTOMATIC START is first page on the menu. This option enables the end user to start furnace automatically at the set value of the date and time.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

This display is simulation of auto start option which indicates set of the date, time and program including heating receipt of end user.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Value between "< >" signs can be changed and values are saved via "ENTER" button. As soon as "< >" signs disappear on the display, furnace is ready to start heating. If 'START' button is pressed, heating program starts to heat up the furnace.

20 SYSTEM STATUS

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

It's an informative display. You can get short info which explains the status and conditions of furnace. You can see sub menus included in SYSTEM STATUS menu respectively as follows;

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Indicating furnace chamber instantaneous temperature and temperature of control PCB.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Menu 22, let's see the date, hour information and enables to change these values.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Furnace language is altered by using "ENTER" button. "▲" ve "▼" buttons enable to change the language and "ENTER" button enables to continue furnace process by set language.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

MENU 24 is to see serial number of furnace and total heating time of the furnace from the first operation of the device.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

MENU 25 indicates average temperature of cumulative working of furnace life. This data is used to determine extra guaranty period.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

DOOR OPEN sensor is optional for BFMF-1200-20 furnaces. In quartz tubes placed resistances are not a cause of short circuit or electrical danger as bare resistances do. If your furnace includes this optional sensor, MENU 26 is displayed in SYSTEM STATUS title. If you set this alarm as 'ON' and if the door is open, heating resistances are shut down. Red heater led is turned off. You are warned by a sound sensor; if the door is open more than 120 seconds. When the door is closed, heating program continuous.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

BFMF-1200-20 Furnaces are calibrated by Martel TC 100 Precision Thermocouple Calibrator. But if you need to calibrate your furnace by a certified calibrator, MENU 27 enables you to correct any temperature diffraction between +- 10oC. And you can keep absolute temperatures which you want to work.

30 PROGRAMS

HEATING PROGRAM-1

Chamber temperature, temperature heating rate, temperature waiting steps, waiting times can be set by this menu option. Programs which are called 40, 50, and 60 are similar options for other program settings.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Values can be set, as it was explained in title; 'OPERATING BY MENU '. 'MOD' is character of heating process that means 'DIRECT' or 'STEPPED' heating. 'RAMP' is to determine the heating rate which specifies the time to let your furnace reach to set maximum temperature of the steps of the heating programs.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Heating rate can be selected between 10°C/min-25°C/min. If you tend to choose higher value than 25°C/min, 'Maximum' appears on the display. On this mode, furnace heating rate is set maximum value automatically.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

"Temp-1" indicates the first step where temperature is kept stable/constant for the stepped program. If ''Direct heating'' is set, 'Temp-1'' is target temperature for a continuous heating without any waiting step. ''Waiting Time'' is time to keep the temperature of furnace stable at the set value. Depending on the set values, furnace heating continues to reach the next target temperature, after the step temperatures time is ended.

If '' Direct Heating'' program is chosen, "Temp-1" is the value to keep the furnace stable at the set temperature, as soon as target temperature is reached for the program. If waiting temperature is tended to be decreased after value 1 is reached, waiting temperature is set as 'Infinite'. In this mode, heating program can be stopped manually.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Menu 33, 34 ve 35 options can be processed, if stepped program is chosen. If you would like to skip any step, step temperature should be set as '0'. For example, if you need to use 3 step program ending at 965°C, you should set your third step temperature as 965°C and fourth step temperature as 0. When the third step is over, fourth step is not considered and all heating program is ended after the third step.

90 ALARM LIST

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

It is informative menu if some failure occurs such as thermocouple failure, PCB overheating.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Temperature sensing system has problem. Most probably thermocouple has a defect. If you have this message, please get in touch with your distributor or TETRA ISI directly.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Control unit protected against overheating by a sensor. If the control unit is overheated, heating power is cut off. This automatic system is directed by control board sensor. You get this message, if this sensor has a defect and if you have this message on the display, please get in touch with your distributor or TETRA ISI directly.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

This message is displayed, if actual time (Date and time) circuit has any problem.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

If the heating chamber exceeds continuous working temperature of the furnace.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

If the heating chamber exceeds maximum temperature of the furnace. Menu 95 is displayed and sound warning starts to warn you.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

If the PCB temperature exceeds 65°C, this message is followed by sound warning and heating power is cut off. When the PCB is cooled down and if 55°C is kept, heating program starts to continue.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

If the furnace is heated up faster than set value, this message is displayed and sound warning starts to work.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

The furnace time is powered by a battery in the control unit. If you see this message on the display, battery should be replaced.

8. START AND STOP PROGRAM

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

You should reach this menu to be able to start heating. If ''ENTER'' button is pressed, 'HEATING' option is activated and heating program selection appears on the display.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

Any heating program can be selected by "▲" ve "▼" buttons and program starts by pressing green color 'START' button and heating is activated.

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating

When heating is activated, the process can be followed by this screen. Continuing program, chamber temperature, waiting time (if the process is at the waiting level) can be observed by this screen.

While furnace is working, program can be stopped by pressing' STOP' button. You should press ' START' to continue Double pressing ' STOP ' lets you come out from the program.

APPENDIX-1 Technical Specifications

1-On / off button, 2-Led indicating on going heating process, 3-Led indicates that furnace works, 4-Display, 5-Heating


BFMF-1300-100 BFMF-1300-20 BFMF-1300-40

BFMF-1300-100 BFMF-1300-20 BFMF-1300-40

BFMF-1300-100 BFMF-1300-20 BFMF-1300-40

APPENDIX-2 Control Unit Features

BFMF-1300-100 BFMF-1300-20 BFMF-1300-40






Request Quote
Name*:
Email*:
I am *: Distributor End User Others
WhatsApp :
Product :
Subject :