Cold Trap Bath BGI2C2 (BBCT-2502)
Cold Trap Bath- Sea, Air, Door to Door Shipping
- 1 Year Warranty
- US & European Standards

Specifications
| Model | BGI2C2 |
| Collection methods | Immersion of Glass Condenser |
| Collection amount | Max. 0.4Kg |
| Lowest temp | -80°C |
| Safety function | Delayed Start of Compressor, Leakage, Overcurrent, Overvoltage Protection |
| Refrigerating capacity | Air Cooling 55W R404A, R23 |
| Cover Interface material | Import PC |
| Tank Interior dimension (mm) | φ220mmx180mm |
| Capacity (L) | 5L |
| PC capping diameter | φ50.3mm 3 holes |
| Condenser diameter | φ10mm (Match the Diaphragm Pump) |
| Interior dimension WxDxH (mm) | 500x600x640 |
| Power | 1300W |
| Electrical requirement | 220V/50Hz |
Description
In vacuum applications,a cold trap is a device that condenses all vapors(except thepermanent gases) into a liquid or solid. The main purpose is to prevent vapors being
produced by an experiment from entering the vacuum pump where they would condense
and contaminate it. Clod traps can also cool surfaces or baffles to prevent oil vapors flowing
from a pump into a chamber. In such a case,a baffle or section of pipe containing a number
of cooled vanes, will be attached to the inlet of an existing pumping system.
Features
Improve pump working efficiency:1. The low temperature of the cold trap can condense the water vapor directly in the cold
trap, thus greatly improving the working efficiency of the vacuum pump.
Protect vacuum pump:
1. Pumps that use oil either as their working fluid (diffusion pumps), or as their lubricant (mechanical rotary pumps), are often the sources of contamination in vacuum systems.
Placing a cold trap at the mouth of such a pump greatly lowers the risk that oil vapors will
back stream into the cavity.
LCD PID controller:
1. P.I.D temperature controller provides accurate and reliable temperature control.
2. Large LCD display screen and interface provides for user-friendly operation.
Energy conservation and environmental protection:
1. Non-freon refrigeration improve cooling efficiency, lower noise, longer life time ensures the stability for long time running.
Safety:
Temperature deviation alarm.
Compressor over current, over heat, over load protection.
Design with
3 glass trap installation.
Upside opened a glass trap.
Easy internal observation with PC Transparent Cover.
Drain valve for easy discharge of collected liquid.
S304 internal bath can be used to do water or ethanol cooling experiments. If equipped
with glass condenser, it also can be used to deal with acid or organic solvents.
Option:
1. RS 485 connector and USB connector can connect computer to save the data.
Cold Trap Diagram:

Operating Manual
Download1. Safety warning
2. Product introduction
2.1 Outline drawing (This drawing is for reference only, subject to the actual product!)
2.2 Overview of structural functions
3. Use of the product
3.1 Preparation before use
3.2 Power on
3.3 Operating steps
3.4 Upper deviation alarm setting
3.5 Check the temperature control accuracy
3.6. Methods to improve temperature control accuracy
4. Maintenance and precautions
5. Appendix
5.1 Main Specifications
5.2 Failure handling methods
Packing list
1. Safety warning
! Dangerous (may result in serious loss of property or casualties)
1. The instrument must be grounded and away from sources of electromagnetic interference (must not be zero line or middle line for the ground).
2. Before use, please confirm the voltage and frequency of the power supply match the product requirements.
3. The product should use a separate power outlet, and confirm the plug, socket, good grounding.
4. Do not allow products in operation without turning off the power switch and unplug the power plug.
5. Do not allow free access or cut short product power line.
6. Do not repair without authorization, the company commissioned by the professionals must be repaired.
! Warning (may cause property damage or personal injury)
1. Must fully read and understand this product manual before operation.
2. When unplugging the power plug, do not drag the power cord directly.
3. One of the following conditions, you must unplug the product power plug:
3.1 When replacing the fuse tube.
3.2 Product failure to be checked repair.
3.3 Long time to stop using the product.
3.4 When moving the product.
! Note (may affect the service life of the product cause it could not work normally)
1. The product should be placed on a hard and firm surface to keep it level.
2. The product should be left some space around.
3. The product must be used under certain conditions of use. (See Chapter III 1.1 ~ 1.4)
4. The product must be used under the correct settings. (See Chapter III for details)
2. Product introduction
2.1 Outline drawing (This drawing is for reference only, subject to the actual product!)

Figure 1

Figure 2
① | Controller |
② | Pressure gauge |
③ | Glass collector |
④ | Switch |
⑤ | BBCT-2501 Power plug/BBCT-2502 power cable |
⑥ | Draining interface |
⑦ | Communication interface (optional) |
Table 1
Glass condenser schematic:

Figure 3
2.2 Overview of structural functions
The cold trap is composed of a controller, a box, a refrigeration system, a glass condenser, and a circulation pump. The controller consists of a temperature controller, a temperature sensor, and the like. The box is made of high-quality thin steel plate, the surface is spray-coated, and the inner tank is made of stainless steel. The refrigeration system consists of a compressor, a condenser, an evaporation coil, and the like.
The equipment relies on the rapid cooling of the refrigeration system to reach the required temperature. The glass condenser series collection device is installed in the liquid tank and used in the vacuum system to complete the reliable operation of the vacuum system.
There is an inlet and outlet interface behind the cold trap. When the liquid outlet interface is not in use, the liquid can be discharged to keep the whole machine clean. The temperature controller is a double-row four-digit display digital controller composed of a CPU processor and peripheral circuits. Pt100 platinum resistance is used as the temperature sensing element, which has the advantages of high control precision and good stability, and alarm functions such as timing and over-temperature alarm protection.
Extended function - optional RS485/232 communication interface, through the connection of the computer, display and monitor the experimental data.
3. Use of the product
3.1 Preparation before use
1.1 The product should work normally under the following conditions of use.
a. Ambient temperature: 5°C~35°C; air relative humidity: ≤85%.
b. Atmospheric pressure: (86~106)KPa.
c. Altitude is not higher than 2000 meters.
d. Power supply: 220V/60HZ.
e. Keep away from heat sources, no corrosive gases around, no strong vibration sources and strong electromagnetic fields.
1.2 The base of the equipment to be placed needs to be flat and composed of non-combustible materials.
1.3 Keep at least 300mm of space around the equipment (front and back ventilation grille), well ventilated.
1.4 This equipment is equipped with a compressor. After the equipment is transported, do not operate immediately. After standing in the correct position, wait for (1~2) days to start working again to ensure that the compressor works normally.
1.5 When filling the bath, please be careful not to pour the liquid into the inside of the meter. It is recommended to open the heat cover to add water or other medium from the sink. The liquid must cover the evaporator.
1.6 Different baths should be added according to different temperature control ranges (super temperature is strictly prohibited).
3.2 Power on

Figure 4
3.2.1.1 Indicator light description
1)TEMP: Display the measured temperature.
2)SET: Display the setting temperature.
3)Cooling light: Light when the cooling output.
4)TIME area: Display the running time or parameter value.
5)Silencer light: There is an alarm, light when the buzzer has silenced.
6)Alarm light: Light when alarm.
3.2.1.2 Key Description
①
Function key: Used for setting value modification; parameter recall and parameter modification confirmation.
②
Shift key: used to set value, shift of internal parameters and view of ambient temperature.
③
Decrease key: In the setting state, click once to decrease 1 digit unit, long press the button, will jump to decrease more digits.
④
Add key: In the setting state, click once to increase 1 digit unit, long press the button, will jump to add more digits.
⑤
Start/stop key: Press and hold this button for more than 4 seconds to control cooling operation / stop.
3.3 Operating steps
1)Timing function: Press
key once, When the time is set as 0, there is no timing function; when the time setting is not 0, the controller has the timing function. Press
key once, the TIME value flashes, indicating that the time can be set as needed, by add, decrease and shift key, set the desired time value. Timing time to, Time window shows "END", buzzer sound, press any key to silence.
2)Temperature setting:Press
key again, "TEMP SET" window flashes, indicating that the temperature can be set as desired. Set the desired temperature by add, decrease and shift key. Then press
key again, return to standard display mode.
Note:①Every time modify a parameter,,need to press"
" key to confirm the modification is valid.
②After all the parameters are set, long press "
" key for about 4 seconds to start running.
3.4 Upper deviation alarm setting
The setting of the upper deviation is reasonable, can play the role of system temperature control or out of control protection, the product must be used when working.
Example: If the product is shipped with AL = 3.0, the alarm temperature is: (set temperature + AL value) °C.
3.5 Check the temperature control accuracy
3.5.1 With 0.1 °C index mercury thermometer (or resolution 0.1 °C digital thermometer) into the bath, the mercury end should be immersed in the bath and placed in the geometric center of the bath.
3.5.2 Choose a point within the range of product control, when the temperature measurement value is equal to the set value, then keep the temperature for about 1 hour. Observe that the difference between the actual measured value of the mercury thermometer and the measured value of the thermometer should be ≤ ± 0.5 °C.
3.6. Methods to improve temperature control accuracy
3.6.1When the product is used for a period of time, check the temperature control accuracy according to 5.2 method. If it exceeds ± 0.5 °C, it can be amended as follows:
3.6.2 Enter the parameter setting, find"
" symbol,

After the formula is calculated, the value of PK is changed on the basis of the original factory (Note: a correction is not allowed, can be repeatedly modified until it meets).
4. Maintenance and precautions
1. The device must be connected to a well-grounded electrical outlet! To ensure personal safety.
2. When the bath is water, please pay attention to the use of temperature to prevent icing, resulting in inaccurate temperature control.
3. When replenishing the bath, the type of bath in the bath must be the same; (please note that the liquid medium in the working chamber must be filled with different bath according to different operating temperatures).
4. Use a flammable bath! Responsible for damage caused by improper use of the bath!
5. Connect the rubber hose, and press the clamp at the joint to prevent it from falling off.
6. Under continuous operation, the connection between the part of the bath cover and the circulation pump will become very cold. Therefore, when touching these places, be careful and prevent frostbite.
7. After filling the tank, carefully transport or tilt it at will, so as to avoid danger or equipment damage caused by immersion of cold liquid into the machine.
8. Must use clean water or medium to strictly prevent foreign matter such as sediment from entering the cooling system and damaging the circulating pump.
9. After use, turn off the power switch, unplug the power cord, insert the drain plug to drain (please pay attention to the bath temperature, be careful! To avoid frostbite) and dry the cabinet.
5. Appendix
5.1 Main Specifications
Cold trap technical parameter table | |||
Product name | Small cold trap device | ||
Type | BBCT-2501 | BBCT-2502 | |
Collection method | Glass condenser immersion | ||
Performance | Collection amount | Max.0.5Kg | Max.0.15Kg |
Lowest temperature | ﹣40°C | ﹣80°C | |
Features | Security function | Compressor delay start, machine leakage, overcurrent, overvoltage protection | |
Composition | Cooling capacity and refrigerant | Air cooling 400W R404A | Air cooling 150W R404A,R23 |
Specification | Cover interface material | Import PC | |
In-tank size(mm)capacity(L) | φ220mm*180mm 6.8L | ||
PC slot cover connection diameter | φ50.0mm 3 hole | ||
Glass condenser connection caliber | φ10mm (Match with diaphragm pump) | ||
Ambient temperature range | 5-35°C | ||
External size(mm) | 315*500*570 | 500*600*710 | |
Power supply, machine power | 220V/60Hz/850W | 220V/60Hz/1300W | |
Way of use | Add ethanol or other antifreeze to the cold trap tank, then immerse the glass condenser in the ethanol liquid, The acid or solvent vapor is condensed and trapped in the glass condenser by cooling and decompression in the glass condenser | ||
Table 2
5.2 Failure handling methods
Fault phenomenon | Reason for failure | Troubleshooting |
1. No supply (Light does not shine) | The plug is not inserted well or the wire is disconnected. | Reinsert or repair |
Power line break or power switch is broken | Replace and repair | |
2.Meter no display | The temperature controller power supply is bad or the incoming terminal is off. | Replace the temperature controller or reconnect the incoming line |
3. Instrument PV screen display ---- | Pt100 sensor is faulty or instrument input circuit is faulty | Replace the Pt100 or replace the temperature controller |
The temperature controller's full scale setting error makes the measured value out of range | Reset | |
4. Not cooling | Refrigeration power switch is not open | Open |
The compressor is broken, the protector is broken, and the boot does not start. | Replace the compressor | |
Overheat causes compressor protector to start | Shut down for several hours, naturally recover | |
Refrigerant leakage, insufficient cooling capacity, blocked pipes | Discharge, add refrigerant | |
Evaporator frosting is serious | Warming | |
5. Large temperature control error | Pt100 sensor has poor contact and large error | Replace Pt100 |
Correct SC, HL parameters (large error) | Refer to this manual | |
Adjust parameters such as PID (static difference) | Refer to this manual | |
6. Poor uniformity | The circulating pump has insufficient flow or the magnetic stirrer is broken, and the bath does not circulate. | Adjust or replace |
Table 3
Packing list
Product name: Cold trap
Serial number | Category | Name | Unit | Quantity | Remark |
1 | Document | Operation instruction | 1 | ||
2 | Document | Packing list | 1 | ||
3 | Spare part | Power line | 1 | BBCT-2502 type no | |
4 | Spare part | Pipe | 2 |