Single Beam UV Visible Spectrophotometer BSSUV-201

Single Beam UV Visible Spectrophotometer BSSUV-201

  • Sea, Air, Door to Door Shipping
  • 1 Year Warranty
  • US & European Standards
Designed to conserve benchspace, our product offers the measurement range of 190-1100nm at bandwidth of 2nm. Engineered for a wide range of applications, our systems provide dependable performance and reliable results. It produces the photometric range of -0.3 to 3 A; 0-200 %T and photometric accuracy of ±0.3%T which makes it suitable for various quantitative assays.
  • Wavelength range: 190nm-1100nm
  • Bandwidth: 2nm
  • Detector: Silicon Photodiode
  • Stray Light: It is achieved in range of ≤0.05 %T
Specification Features ApplicationsAccessories Optional Accessories
Wavelength Range190-1100 nm
Wavelength Accuracy±1nm
Wavelength Reproducibility≤0.3 nm
Wavelength Repeatability≤0.3 nm
Spectral Bandwidth2 nm
Photometric Range0.200%T, -0.3-3A, 0-9999C
Photometric Accuracy±0.5 %T
Photometric Repeatability≤0.2 %T
Stability±0.0001 A/h at 500 nm
Baseline Flatness±0.004 A
Stray Light≤0.05 %T
Light SourceDeuterium Lamp and Tungsten-Halogen Lamp
Gross Dimension (W/D/H)420x300x160 mm
Display128x64 Dots LCD
Noise±0.004 A
Weight (Net/Gross)13 kg
Power300 W
Power SupplyAC 220V/50Hz or AC110/60Hz
  • Wavelength range: 190nm-1100nm
  • Bandwidth: 2nm
  • Detector: Silicon Photodiode
  • Stray Light: It is achieved in range of ≤0.05 %T
  • Light source: Deuterium and tungsten halogen lamp
  • A wide range of standard cuvettes can be used based on the sensitivity or sample volume requirements
  • The sample analysis software provides the scanning, fixed wavelength analysis, quantitative analysis, data collection, storage, export, and reporting.
  • USB port and parallel port
  • Inclusive of standard PC software
  • With additional feature of an easy access USB port available in the unit which enables results to be stored directly to a USB memory stick for easy transfer of data. It can store up to 200 results.
  • Provides Photometric, Quantitative, Kinetics, Wavelength scan, Multi-wavelength and DNA/Protein Test functions
  • Enables building of calibration curves and further implementation of associated tests
  • Formation of multi-point standard curve using calibration curve method enables measurement of unknown sample concentration
  • Implements sample measurement after addition of coefficient of curvilinear equation using coefficient method
  • Automatic wavelength calibration and automatic deviation repair
  • Easy changing of deuterium and tungsten lamp excluding the need for debugging
Most suitable for quantitative determination assays and toxicological assays in environmental, industrial, pharmaceutical fields.
Accessory CodeNameUnit
3000606019Glass cuvette (10 mm)4 pcs
3000606021Quartz cuvette (10 mm)2 pcs
3000606023Dustproof cover1 pc
3001715006Power Cord (US/EU)2pcs
3000606026Software CD1 pc
3000606028USB Cable1 pc
Accessory CodeNameDescriptionLight Path LxWxH (mm)VolumeUnitLight Path LxWxH (mm)
3000606007Single Position Water Jacketed Ambient Cell Holder10 mm
30006060084-Position Cell Holder10 mm
30006060094-Position Cell Holder50 mm
30006060104-Position Cell Holder100 mm
2902406006Adjustable Cell Holder for Micro Cuvettes
2902406007Transparent Film Holder
3000606012Tube Holder
3000606013Thermostatic Controller
3000606014Auto Sampler
3000606015Auto Thermostatic Sampler
3000606016Micro Printer
3000606017Tungsten Lamp 6V; 10W (PH)
3000606018Tungsten Lamp 12V; 20W (PH)
3000606020Tungsten Lamp 12V; 20W (OSRAM)
3000606022D2 Lamp A510YXU Milasi
3000606024Glass Cuvette5x10x451.7 mlSet (4 pcs)
3000606025Glass Cuvette10x10x453.5 mlSet (4 pcs)
3000606027Glass Cuvette20x10x457 mlSet (4 pcs)
3000606029Glass Cuvette30x10x4510.5 mlSet (4 pcs)
3000606031Glass Cuvette40x10x4514 mlSet (4 pcs)
3000606033Glass Cuvette50x10x4517.5 mlSet (4 pcs)
3000606034Glass Cuvette100x10x4535 mlSet (4 pcs)
3000606035Quartz Cuvette5x10x451.7 mlSet (2 pcs)
3000606036Quartz Cuvette10x10x453.5 mlSet (2 pcs)
3000606037Quartz Cuvette20x10x457 mlSet (2 pcs)
3000606038Quartz Cuvette30x10x4510.5 mlSet (2 pcs)
3000606039Quartz Cuvette40x10x4514 mlSet (2 pcs)
3000606040Quartz Cuvette50x10x4517.5 mlSet (2 pcs)
3000606041Quartz Cuvette100x10x4535 mlSet (2 pcs)
3000606042Micro Quartz Cuvette10x2x2.550 mlSet (2 pcs)
3000606043Micro Quartz Cuvette10x2x5100 μlSet (2 pcs)
3000606044Micro Quartz Cuvette10x2x10200 μlSet (2 pcs)
3000606045Micro Quartz Cuvette10x2x25500 μlSet (2 pcs)
3000606046Micro Quartz Cuvette10x2x45700 μlSet (2 pcs)
3001906006Quartz Flow CuvetteSet (2 pcs)
3000606047Micro Quartz CuvetteSet (2 pcs)10x2x45
Request Quote
Name*:
Email*:
I am *: Distributor End User Others
WhatsApp :
Product :
Subject :
Message :

Operating Manual for BSSUV-201

Download Operating Manual PDF

CHAPTER 1 PREFACE
    1. Safety
    2. General
    3. Electrical
    4. Warning
    5. Performance
    6. Radio Interference
CHAPTER 2 INTRODUCTION OF THE INSTRUMENT
    1. Brief Introduction of the Instrument
    2. Working Principle
    3. Structure
    4. Unpacking Instructions
    5. Specifications
CHAPTER 3 INSTALL INSTRUMENT
    1.Environment Required
    2. Installation
CHAPTER 4 OPERATION INTRODUCTION
    1. Software Structure
    2. Basic Operation
    3. Preparation before test
    4. Measure
    5. System Functions
CHAPTER 5 Instrument Maintenance
    1. Daily Maintain
    2. Troubleshooting
    3. Fuse Replacement

CHAPTER 1 PREFACE

1. Safety

Read the following before installing and using the instrument and its accessories. This instrument should be operated by appropriate laboratory technicians.

2. General

The apparatus described in this manual is designed to be used by properly trained personnel in a suitable equipped laboratory. For the correct and safe use of this apparatus it is essential that laboratory personnel follow generally accepted safe procedures in addition to the safety precautions called for in this manual.
The covers on this instrument may be removed for servicing. However, the inside of the power supply unit is a hazardous area and its cover should not be removed under any circumstances. There are no serviceable components inside this power supply unit. For this instrument, avoid touching the high voltage power supply at all times.
Some of the chemicals used in spectrophotometry are corrosive and/or inflammable and samples may be radioactive, toxic, or potentially infective. Care should be taken to follow the normal laboratory procedures for handling chemicals and samples.

3. Electrical

Before switching on the apparatus, make sure it is set to the voltage of the local power supply (see Installation).
The power cord shall be inserted in a socket provided with a protective earth contact. The protective action must not be negated by the use of an extension cord without a protective conductor.

4. Warning

Any interruption of the protective conductor inside or outside the apparatus or disconnection of the protective earth terminal is likely to make the apparatus dangerous. Intentional interruption is prohibited.
Whenever it is likely that the protection has been impaired, the apparatus shall be made inoperative and be secured against any unintended operation.
Note:NEVER touch or handle the power supply on this instrument due to the high voltage! The protection is likely to be impaired if, for example, the apparatus
▲Shows visible damage
▲Fails to perform the intended measurements
▲Has been subjected to prolonged storage under unfavorable conditions
▲Has been subjected to severe transport stresses

5. Performance

To ensure that the instrument is working within its specification, especially when making measurements of an important nature , carry out performance checks with particular reference to wavelength and absorbance accuracy. Performance checks are detailed in this manual.

6. Radio Interference

For compliance with the EMC standards referred to in the EC Declaration of Conformity, it is necessary that only shielded cables supplied by us are used when connecting the instrument to computers and accessories.

CHAPTER 2 INTRODUCTION OF THE INSTRUMENT

1. Brief Introduction of the Instrument

This instrument is a single beam, general purpose instrument designed to meet the needs of the Conventional Laboratory, This instrument is ideal for various applications, such as: Chemistry, Biochemistry, Petro-chemistry, Environmental Protection, Food and Beverage Labs, Water and Waste Water Labs and other fields of quality control and research.
This instrument incorporates a 128×64 dots matrix LCD display for photometric results, easy operation and wavelength range of 200nm to 1000nm. This instrument is ideal for measurements in the visible and ultraviolet wavelength region of the electromagnetic spectrum.


Fig 2-1 Front View



Fig 2-2 Control Panel

2. Working Principle

Different matter has different but special absorbance wavelength point. Also, when at the fixed wavelength point, the absorbance has some relation to the substance’s (Always transparent Solution) concentration and its thickness. The relation can be concluded as the following Formula which is called Lambert-Beer Law.
T= I/Io
A=KCL= -log I/Io
A Absorbance
C Concentration of the Solution
K Absorbance Coefficient of the Solution
L The length of the Solution in the light path
I The intensity of the light focused on the A/D after it permeate the solution to be measured. I o The intensity of the light focused on the A/D after it permeate the Solution.
Note: When test, the solvent is usually taken as the Reference Solution and its Transmittance is considered as 100%T. While the Transmittance of the sample to be tested is a relative value which is got comparing to that of the Reference.

3. Structure

The spectrophotometer consists of five parts:
1. Halogen or deuterium lamps to supply the light;
2. A Monochromator to isolate the wavelength of interest and eliminate the unwanted second order radiation;
3. A sample compartment to accommodate the sample solution;
4. Detector to receive the transmitted light and convert it to an electrical signal; and
5. A digital display to indicate absorbance or transmittance. The block diagram (Fig 2.4) below illustrates the relationship between these parts.


Fig 2.4 Block diagram for the Spectrophotometer

In your spectrophotometer, light from the lamp is focused on the entrance slit of the mono-chromator where the collimating mirror directs the beam onto the grating. The grating disperses the light beam to produce the spectrum, a portion of which is focused on the exit slit of the mono-chromator by a collimating mirror. From here the beam is passed to a sample compartment through one of the filters, which helps to eliminate unwanted second order radiation from the diffraction grating. Upon leaving the sample compartment, the beam is passed to the silicon photodiode detector and causes the detector to produce an electrical signal that is displayed on the digital display.

4. Unpacking Instructions

Carefully unpack the contents and check the materials against the following packing list to ensure that you have received everything in good condition.
Packing List
—--------------------------------------------------------------------------------Description Quantity
Spectrophotometer……………………………………………………………. 1
Power Cord……………………………………………………………………… 1
1cm Glass Cuvettes…………………………………..………………………….1 Set of 4 1cm Quartz Cuvettes…………………………………...………………. ………1 Set of 2 User’s Manual…………………………………………………………………… 1
Dust Cover……………………………………………………………………… 1

Note: The Printer mentioned in this manual is an optional accessory, it does not come standard with the instrument.

5. Specifications


Wavelength Range190-1000nm
Band Width2nm
Wavelength Accuracy±1nm
Wavelength Repeatability0.5nm
Wavelength SettingAuto
Photometric Accuracy±0.5%T
Photometric Repeatability0.3%T
Photometric Display Range0-200%T, -0.3-3.0A, 0-9999C
Stability0.002A/h @ 500nm
Stray Light≤0.3%T@220nm, 360nm
Data Output PortUSB
Printer PortParallel Port
Display128*64 Dots LCD
LampsDeuterium Lamp & Tungsten Halogen Lamp
DetectorSilicon Photodiode
Power RequirementAC85~250V
Dimension (L*W*H)420*280*180mm
Weight11kg

CHAPTER 3 INSTALL INSTRUMENT

1.Environment Required

To ensure the best performance, the following conditions are required:
- The best working temperature range is 16-30℃ and the humidity is 45-80%.
- Keep it as far as possible away from the strong magnetic or electrical fields or any electrical device that may generate high-frequency fields.
- Set the unit up in an area that is free of dust, corrosive gases and strong vibrations.
- Remove any obstructions or materials that could hinder the flow of air under and around the instrument.
- The power requirement is AC85V~265V.
- Use the appropriate power cord and plug into a grounded outlet.
- If the local voltage is not stable enough, a Power Stabilizer is required.
- Be away from direct sunlight

2. Installation

Step 1: Check the packing list
Unpack the contents, check the materials with the Packing List. Any damage or Lost found, please contact us or the local dealer.
Step 2: Position
Place the instrument on the stable table carefully.
Step 3: Install printer (Optional)
Make sure the printer is power off, link the printer’s data cable to the instrument’s parallel port.
Step 4: Link the power cord
Make sure the instrument’s power switch is in the Off condition, link the power cord to the instrument and insert another end in the socket provided with a protective earth contact.
Step 5: Switch on the power
Check again, make sure that all the links are right. Switch on the power. The instrument can be used to measure after self-test and pre-warm.

CHAPTER 4 OPERATION INTRODUCTION

1. Software Structure

1.1 Main Interface


Fig 4-1 Main Interface

After 20 minutes’ warm-up, the system goes into main interface. (Fig.4-1) Press the key of to change the functions.

1.2 Firmware Functions
The firmware consists of 5 functions: T, A, C, F and system functions
1) T
To measure the Transmittance of a sample in a fixed wavelength point. The test result can be stored in the RAM, total 200 groups of data can be stored.
2) A
To measure the Absorbance of a sample in a fixed wavelength point. The test result can be stored in the RAM, total 200 groups of data can be stored.
3) C-Standard Curve Method
Set up standard curve by standard samples; use the new curve to measure the unknown sample’s concentration. The curves and the test results can be stored in the RAM. Total 200 curves can be stored.
4) Coefficient Method
Input the values of the curve equation’s coefficient, and then measure the unknown sample’s concentration.
5) Setup-System Functions
Press the key of on the keypad to go into system functions (Only available in T MODE and A MODE). Light source management; Get Dark Current; Calibrate Wavelength and Load default, etc.

2. Basic Operation

2.1 Set Wavelength
In the measurement to go into wavelength setting Interface. Use Arrow Key to increase or decrease the value of wavelength,then to confirm. When arriving at the setting wavelength point, it sets 100%T/0Abs automatically.
2.2 Delete the test result and stored data
In the measurement Interface,press to delete the test result and stored data.
2.3 Calibrate 100%T/0Abs
Put the Reference in the light path, press to calibrate 100%T/0Abs。
2.4 Text BoxMeasure the samples
Put the samples in the light path and press Print the test result to measure.
2.5 Print the test result
In the Measurement Interface, press to print the result.

3. Preparation before test

3.1 Switch on the instrument and begin the system self-test(Diagnostic Program)
Make sure to remove all the blocks in the light path; Close the compartment lid and switch on the power; Then the system will begin self-test.
Note: During the course of Diagnostic Program runs, please DON’T open the lid of the compartment!
3.2 Pre-warm
When the diagnostic program finished, it goes into pre-warm condition. 20 minutes is required before measure.
3.3 Check the Cuvettes
The cuvette should be clear and there’s no solution remains on the transparent surface. If the wavelength is below 340nm, Quartz Cuvette is a must.

4. Measure


4.1 T-Mode
Step 1, Go into T Mode Interface
In the Main Interface , Key to move the cursor on “T” , you’ll see the current sample’s T% value. Press to go into Continuous Measurement Interface.(Fig. 4-2)

Fig. 4-2

Step 2, Set Wavelength
Press to set Wavelength,input the wavelength value by the Arrow Key, then press to confirm.(Fig. 4-3)


Fig. 4-3

Step 3, Calibrate 100%T/0Abs Put the Reference Solution in the light path,press to calibrate 100%T/0Abs;
Step 4, Measure Sample
Put the unknown concentration sample in the light path, press to measure,and the result displays in the data sheet. At the same time, the test result will be stored in the RAM automatically.
Repeat this step to finish all the samples measurement.
Step 5, Print or Delete the data
Press to go into the following interface.(Fig.4-4)

Fig.4-4

1. When the optional Printer is linked to the instrument, if you want to print out the test result, use Arrow Key to choose “Print”, then press to confirm. After printing, the data will be deleted from the RAM
2. If you want to delete the saved data, move the cursor on the item of “Delete”, then press to confirm.
4.2 A-MODE
Step 1, Go into A Mode Interface
In the Main Interface, press Key to move the cursor on “A” , you’ll see the current sample’s Absorbance value.
Press to go into Continuous Measurement Interface.(Fig. 4-5)


Fig.4-5

Step 2-5 is the same as T-mode, please refer to the corresponding operation in 4.1.
4.3 C-MODE (Standard Curve)
1. Go into C-MODE Interface
In the Main Interface, press Arrow Key to move the cursor on “C”, then you’ll see the following Interface. (Fig.4-6) Here you can create a new curve, or you can load a saved curve. If you want to delete a saved curve, just choose “Delete Curve”.

Fig. 4-6

2. Create Curve
If you want to create a new standard curve by several standard samples, Use Arrow Key to move the cursor on “Create
Text BoxCurve”, then press to confirm. The system will ask you to input the quantity of your standard samples. (Fig.4-7)

Fig.4-7

Note: The default quantity is “3”. Please don’t change the value at this time, you should set the Wavelength at first.
Step 1, Set Wavelength

Fig. 4-8
Press to go into the wavelength setting interface,input the WL. Value by the Arrow key and press to confirm.(Fig. 4-8)
Step 2, Input the quantity of the standard samples
Use Arrow Key to input the quantity of the standard samples, then press to confirm. The system will go into the following interface. (Fig. 4-9)

Fig. 4-9

Step 3, Calibrate 100%T/0Abs
Put the Reference Solution in the light path,then press to calibrate 100%T/0Abs

Step 4, Input Standard Samples’Concentration

- Pull the 1# Standard Sample in the light path and input the value of its concentration by the Arrow key, then press to confirm. The system will go into the next interface. (Fig.4-10)
- Pull the 2# Standard Sample in the light path and input its concentration, then press
to confirm.
- Repeat the above step till the last sample’s concentration is finished.

Step 5, Display the Curve
When the last standard sample finished, double press to confirm. then the Standard Curve and its equation will display on the screen automatically. (Fig. 4-11) At the same time, the Curve Equation will be saved in the RAM automatically.

Fig. 4-11
Note: If there’s some mistake during the operation, the system will beep 3 times and return the initial interface automatically. Of course, the curve can not be displayed on the screen.

Step 6, Measure Samples

Fig. 4-12- Pull the Reference solution in the light path and press to go into the continuous measurement interface. (Fig. 4-12)
- Pull the unknown sample in the light path, then press Text Boxto measure.
The result will be displayed on the screen one by one. At the same time, the data will be saved in the RAM automatically. Total 200 groups of data can be saved.

3. Load Curve
All the curve equations will be saved in the RAM automatically. If you want to reload the saved curve, just follow the following steps:

Fig. 4-13

- Move the cursor on “Load Curve” and press to confirm. Then the system will go into the curve equation record interface.(Fig. 4-13)


Fig. 4-14
Move the cursor on the equation that you need and press to confirm. Then the corresponding curve will display on the screen. (Fig. 4-14)

Press to go into sample continuous measure interface and measure samples. (Fig. 4-12)
Note: When you reload the curves, the wavelength will be set at the point where you creat the Standard Curve automatically.

4. Delete Curve

Fig. 4-15
Move the cursor on “Delete Curve” and press to confirm. Then the system goes into the curve equation choosing interface. (Fig. 4-15)
- Move the cursor on the curve equation that you want to delete and press , then system will ask you to confirm again. (Fig. 4-16)

Fig. 4-16

- Move the cursor on “Yes” and press to confirm, then the curve equation will be deleted. If you don’t want to delete it, just choose “ ” or press to return.
4.4 F-MODE (Coefficient Method)
1) Go into F-MODE Interface

Fig. 4-17
Press Arrow Key to choose “F” in the Main Interface, and you’ll see the following Interface. (Fig. 4-17)
2) Parameter Setting
- Use Arrow Key to move the cursor on “Coefficient K” and press to go into K setting interface.Fig. 4-18;Input every bit of K by Arrow Key and press to confirm. When the last bit finished, system will return the former Interface automatically.

Fig. 4-18
- Input the value of B in the same way(Fig. 4-19)

Fig.4-19
3) Test Samples
Step 1, Go into pre-test Interface
Move the cursor on “Test”, then press to go into pre-test Interface. (Fig.4-20)

Fig.4-20
Step 2, Set Wavelength

Fig. 4-21
Press to go into wavelength setting interface,input the wavelength value by Arrow key and press to confirm. (Fig. 4-21)
Step 3, Calibrate 100%T/0Abs
Press to go into continuous measurement interface (Fig. 4-22). Pull the Reference Solution in the light path,then press to calibrate 100%T/0Abs;

Fig. 4-22
Step 4, Measurement
Pull the samples to be measured in the light path, press to measure, then the result will be displayed on the screen . Repeat this step to measure other samples. All the test results will be listed on the screen and stored in the RAM automatically. They can be reload even after a sudden power cut.
Step 5, Print or Delete the Test Result
In the continuous measurement interface, press In the T mode or A mode of Main Interface, press functions(Fig. 4-23),4.1 for details.

5. System Functions

In the T mode or A mode of Main Interface, press functions(Fig. 4-23)to print or delete the test result. Please refer to Chapter Key to go into system

Fig. 4-23

1) Energy
To see the energy of present wavelength. Move the cursor on “Energy”, then press Text Boxto confirm. (Fig.4-24)
Note: Users seldom use this function. It’s useful for service man to judge the fault.

Fig. 4-24
2) D2 lamp management
When the wavelength point needed is in the range of 340-1100nm , the D2 lamp can be switched off to prolong its life time.

Use Arrow key to choose “D2 Lamp On” and press to go into D2 lamp setting interface(Fig. 4-25). Use Arrow key to choose “On ”or “Off” and press to switch on or switch off the D2 lamp. Press to return.


Fig. 4-25
3. Get Dark Current
When the working condition changes, A dark current calibration is required before any measurement. Use Arrow key to choose “Dark Current” and press to confirm. (Fig. 4-26)
Note: Remove the cuvettes from the light path before this action, don’t open the lid of the compartment during the course of calibration.
Press to return when it finishes.

Fig. 4-26
When it finishes, it will display the following words on the screen. (Fig.4-27) If it displays “Dark Current Err”, you should check if there’s something else in the compartment and repeat this step.

Fig. 4-27
4. Wavelength Calibration
After some time, the energy of lights would drop off and there will be some but slight influence on the test result. In this case, users can reset the wavelength to compensate. We suggest the users to calibrate the wavelength every one or two months.
Remove all the blocks in the light path and close the lid of the compartment, press Arrow Key to choose “Calibrate WL.” and press to begin the calibration ( Fig. 4-28 ) . The lid of the compartment should not be opened during the calibration.

Fig. 4-28

Fig. 4-29

When it finishes, it will display the following words on the screen. (Fig.4-29) If it displays “Wavelength Err”, you should check if there’s something else in the compartment and repeat this step.
5. Load Default
Press Arrow Key to choose “Load Default” and press to confirm.(Fig. 4-30), press Arrow Key to choose “Yes” and press to confirm. Then the system will recover the default setting when leave factory. Choose “No” with pressed to cancel the operation and return.

Fig. 4-30
Note: All the saved data and curves would be deleted from the RAM once “Yes” is chosen!
6. Version Press Arrow Key to choose “Version” and press to confirm. You’ll see the version of the software and hardware. (Fig.4-31). Press to return.

Fig. 4-31
Note: For the reason of update, the version you see on the screen may be different from the above one.

CHAPTER 5 Instrument Maintenance

To keep the instrument work in good condition, constant maintain is needed.

1. Daily Maintain

1) Check the compartment
After measurement, the cuvettes with sample solutions should be taken out of the compartment in time. Or the volatilization of the solution would make the mirror go moldy. Users must pay more attention to the corrosive sample and liquid easy to volatilize. Any solution remains in the compartment should be wiped off immediately.
Surface Clean
2) The cover of the instrument is with paint. Please use wet towel to wipe off the drip on the surface immediately. Organic solution is forbidden to be used to clean the cover. Please wipe off the dirt on the cover timely.
3) Clean the cuvettes
After every test or after a solution change, the cuvettes should be cleaned carefully, or the remains on the surface would cause measuring error.

2. Troubleshooting

1) Dark Current Error when self-test
Possible Cause
Open the lid of the compartment during the course of self-test
Solution
Close the lid of the compartment and switch on the power again.
2) No response after power on
Possible Cause
- Bad contact in power supply
- Fuse melt
Solution
- Improve the contact
- Replace a new fuse
3) Printer not work, printer error
Possible Cause
- No power supply
- Bad contact in power supply
- Bad contact of the data cable
Solution
- Switch on the power supply
- Improve the contact
- Improve the contact
4) No Stable Reading
Possible Cause
- No enough pre-warm
- Glass cuvettes used in UV Range
- No stable Sample
- Much higher sample concentration
- Low voltage or unstable power supply - Lights defect
- Light used up
Solution
- Increase the pre-warm time
- Use Silicon Cuvettes.
- Improve the sample
- Dilute the sample
- Improve the power condition
- Replace a new lamp
- Replace a new lamp
5) Worse Repeatability
Possible Reason
- Unstable sample
- Cuvettes polluted
Solution
- Improve the sample
- Clean the cuvettes
6) Incorrect reading
Possible Reason
- Dark Current Error
- Worse matching of the cuvettes
Solution
- Re-get the dark current
- Improve the matching of the cuvettes

3. Fuse Replacement

Danger! Be sure to switch off the power and unplug the socket before replacement!
Step 1: Tools preparation
Prepare a 3×75 Flat Blade screwdriver
Step2:Switch Off the power supply
Switch off the power supply, and unplug the socket.
Step 3: Take out the Fuse Seat
Take out the fuse seat by the screwdriver. (Fig.5-1)

Fig. 5-1
Step 4: Replace a new fuse
Pick out the spare fuse and replace it to the working position. (Fig.5-2)

Fig. 5-2
Step 5: Reset the fuse seat
Replace the fuse seat in the power socket
Step 6: Switch on the power
Plug the socket and switch on the power