Multi-parameter Analyzer BMET-602

Multi-parameter Analyzer BMET-602

  • Sea, Air, Door to Door Shipping
  • 1 Year Warranty
  • US & European Standards
Multi-parameter Analyzer simultaneously measures Ion, Conductivity, Dissolved Oxygen, Temperature that otherwise requires four different meters. With the appropriate probes installed, it can also simultaneously measure several or one parameters.
    Specification Features
    pH: 
          Range-3.00 to 20.00 pH
          Resolution1.1, 0.01 pH
          Accuracy± 0.01 pH
          Calibration PointsUp to 6
          Standard CustomizationYes
          Standard RecognitionNIST, GB and DIN buffers
    mV: 
          Range-2001.00 to 2000.00 mV
          Resolution1.1 mV
          Accuracy± 0.03 mV or ± 0.1 % of reading whichever is greater
    pX: 
          Range- 2.00 to 20.01
          Resolution1.1, 0.01 pX
          Accuracy±0.02 pX
          Calibration PointsUp to 6
    ISE: 
          Range1E-9 to 9.999E10
          Unitmol/L, mmol/L, g/L, mg/L, μg/L, ppm
          ResolutionUp to 4 significant digits
          Accuracy± 0.5 %
          Calibration PointsUp to 6
    Conductivity: 
          Range1.000 μS/cm to 1000 mS/cm
          Resolution1.01Ωcm minimum
          Reference Temperature21, 25 °C
          Calibration PointsUp to 4
          Standard Recognition85 μS/cm, 1413 μS/cm; 12.88 mS/cm
    Resistivity: 
          Range6.00 Ωcm ~20.00 MΩcm
          Resolution1.01Ωcm minimum
          Accuracy± 1.0 % FS
    TDS: 
          Range1.00 mg/L ~300 g/L
          Resolution1.01 mg/L minimum; changed with range
          Accuracy± 1.0 % FS
    Salinity: 
          Range(0.00 ~8.00) %
          Resolution1.01 %
          Accuracy± 0.2 %
    Temperature: 
          Range- 6 to 110 °C, 23 to 230 °F
          Unit°C, °F
          Resolution1.1
          Accuracy± 0.3
    Measurement: 
          Reading ModeAutoRead( Fast, Medium,Slow), Timed, Continuous
          Reading PromptsReading, Stable, Locked
          Temp. CompensationATC, MTC
    Data Management: 
          Data Storage501 results each
          GLP FeaturesYes
    Inputs: 
          pH ElectrodeBNC(Q9)
          Temp./DO Probe5-pin aviation connector
          Temp./EC Probe6-pin aviation connector
    Outputs: 
          USBPC
          USB, RS 233printer
    Display Options: 
          BacklightYes
          Auto Shut-down2~60 min, off
          IP RatingIP55
    General: 
          PowerAC Adapter, 100-240 V AC input, DC9V output
          Dimensions243x195x68 mm
          Weight901 g(1.98 lb)
    Request Quote
    Name*:
    Email*:
    I am *: Distributor End User Others
    WhatsApp :
    Product :
    Subject :
    Message :

    Operating Manual for BMET-602

    Download Operating Manual PDF

    1.Introduction
        1.1. Introduction
        1.2. Technical Specification
        1.3. Function Introduction
        1.4. Manual Introduction
    2. Safety Notices
    3. Terms Explanation
    4. Overview and Installation
        4.1. Overview
        4.2.Instrument Installation
    5. Instrument Operation
        5.1 Switch On/Off
        5.2 Screen Icons
        5.3 Function Key
        5.4 Parameter Settings
        5.5 pH Measurement
        5.6. Ion Measurement
        5.7 Conductivity Measurement
        5.8 TDS Measurements
        5.9. Salinity Measurement
        5.10 Resistivity Measurement
        5.11 Dissolved Oxygen Measurement
        5.12 Data Management
    6. Maintenance/Troubleshooting
        6.1 Meter Maintenance
        6.2 Electrodes Maintenance
        6.3 Troubleshooting
    7. Technical Supports
        Accessories
    8. Appendixes
        Appendix 1
        Appendix 2

    1.Introduction

    1.1. Introduction

    The BMET-602 series benchtop multi-parameter analyzer is a multifunctional water quality testing device equipped with pH/ISE conductivity, dissolved oxygen, temperature modules. The meter supports measurement parameters include pH, mV, ORP, pX, ion concentration, conductivity, resistivity, TDS, salinity, dissolved oxygen concentration, dissolved oxygen saturation, temperature.
    Modules and measurement parameters

    ModuleParametersBMET-602BMET-601
    pHpH
    mV ORP
    ISEpX
    ion concentration
    ConductivityResistivity conductivity TDS
    salinity ash
    Dissolved oxygenDissolved oxygen current dissolved oxygen concentration
    saturation
    TemperatureTemperature

    Table 1
    General Features
    High-resolution LCD display, 5.7 inches. Multi-reading feature allows auto-read, timed-read and continuous-read. Automatic/Manual temperature compensation ensures accurate results. Auto-hold feature senses and locks the measurement endpoint. Data Storage 500 sets (GLP-compliant). Support for USB or RS-232 communication. Reset feature automatically resumes all settings back to factory default options. IP54 waterproof.
    pH
    1-5 points calibration with Standard Recognition. Selectable pH buffer groups, including NIST, DIN, GB, USA. Automatic electrode diagnosis with pH slope and offset display.
    Ion
    1-5 points calibration. Selectable measurement unit, including μg/L, mg/L, g/L, mmol/L, PX, etc. Multi-measurement modes are supported, including Direct Reading mode, Standard Addition mode, Sample Addition mode and GRAN mode. Over 10 methods are built-in, including F-, Cl-, Br-, I-, NO3-, BF4-,NH4+, K+, Na+, Ca2+, Cu2+, Pb2+, Ag+ and etc., user-defined method is supported.
    Conductivity
    1-3 points calibration with Standard Recognition. Settable parameters, including cell constant, temperature compensation coefficient and TDS factor. Temperature compensation type (none, linear, pure water). DO Support for air-saturated water or zero oxygen calibration. Auto barometric pressure compensation Manual Salinity Factor Correction Selectable pressure unit, including kPa, mbar, Torr, Atm.

    1.2. Technical Specification

    Instrument Specifications

    ModelBMET-602BMET-601
    mVRange(-2000.0~2000.0)mV
    Resolution0.1mV
    Accuracy± 0.1% or ±0.3mV
    Repeatability0.5mV
    Input Current≤1×10-12A
    Input Impedance≥1×1012Oh
    pHRange(-2.00~20.00)pH
    Resolution0.01pH
    Accuracy±0.01pH
    Repeatability0.005pH
    Measurement Accuracy±0.02pH
    Measurement Repeatability0.01pH
    pXRange(-2.00~20.00)pX
    Resolution0.01pX
    Accuracy±0.01pX
    Repeatability0.005pX
    Measurement Accuracy±0.02pX
    Measurement Repeatability0.01pX
    Ion concentrationRange( 1.000e -9 ~9.999e +9 ), mol/L,
    mmol/L, g/L, mg/L, μg/L, ppm, ppb

    Table 2

    ModelBMET-602BMET-601
    Resolution4 significant digits
    Measurement Accuracy±0.5%
    ConductivityRange0.000 μS/cm~1000mS/cm Auto-ranging:
    (0.000~1.999)μS/cm (2.00~19.99)μS/cm (20.0~199.9)μS/cm
    (200~1999)μS/cm (2.00~19.99)mS/cm (20.0~199.9)mS/cm
    (200~1000)mS/cm
    Resolution0.001μS/cm, automatic switching according to the
    range
    Accuracy± 1.0%(FS)
    Repeatability0.33%(FS)
    Measurement Accuracy±1.50%(FS)
    Measurement Repeatability0.50%(FS)
    ResistivityRange5.00Ω·cm~20.00MΩ·cm
    Resolution0.01Ω·cm, automatic switching according to the range
    Accuracy±1.0%(FS)
    TDSRange0.00mg/L~300g/L
    Resolution0.001mg/L, automatic switching according to the
    range
    Accuracy±1.0%(FS)
    SalinityRange(0.00~8.00)%
    Resolution0.01%
    Accuracy±0.2%

    Table 3

    ModelBMET-602BMET-601
    Measurement Accuracy±0.4%
    Dissolved OxygenRange(0.00~20.00)ppm
    Resolution0.01 ppm
    Dissolved OxygenAccuracy±0.10 ppm
    Repeatability0.15 ppm
    Accuracy at 0% DO≤0.1 ppm
    Measurement Accuracy±0.30 ppm
    Dissolved OxygenInstrument indication error±0.10 ppm
    Response time≤ 45s (90% response at 20°C).
    Salinity Compensation
    Accuracy
    ±2%
    SaturationRange(0.0~200.0)%
    Resolution0.1%
    Accuracy±2.0%
    Measurement Accuracy±10.0%
    TemperatureRange(-5.0~110.0)℃
    Resolution0.1 ℃
    Accuracy±0.2 ℃
    Instrument indication error±0.4℃ (0℃~60℃),
    ± 1.0 °C (Else).
    Work environmentAmbient temperature: (0 ~ 40) °C Relative humidity: not more than 85%
    Dimensions (L×B×H), weight (kg).242mm×195mm×68mm,0.9kg
    Power supplyPower adapter, Input: AC100~240V Output: DC9V

    Table 4

    1.3. Function Introduction

    Functions Specification

    FeaturesExplanation
    Basic FunctionBacklight adjustment
    Automatic diagnostics
    Factory reset
    Default parameter
    Prompt Sound
    Time settings
    Power failure protection
    Firmware upgrade
    Anti-interference automatic recovery
    Automatic shutdown
    ProtectionIP54
    Reading FunctionReading stability settings
    Auto-lock reading
    Reading Mode
    Sample ID
    Data ManagementStorage500 sets of measurement parameters each
    View
    Delete
    GLP

    Table 5

    FeaturesExplanation
    Communications and external devicesPrinterRS232 Serial Printer
    Content and format customization
    PC
    pH/mV MeasurementpH electrode status/performanceSlope, Electrode status
    Multi-point calibration5 points
    Automatic standard solutions
    recognition
    4 groups
    Standards customization
    Standard groups customization1 group
    Automatic temperature compensation
    Manual temperature compensation
    pH electrode diagnostics
    MeasurementMulti-point calibration5 points
    Optional units
    Measurement modeDirect reading method, standard addition method,
    sample addition method, GRAN method
    Built-in ions methods
    Ion’s customization
    Conductivity MeasurementConductivity
    Resistivity
    TDS
    Salinity

    Table 6

    FeaturesExplanation
    Conductivity MeasurementConductivity-ash
    Reference temperature20.0°C, 25.0°C
    Multi-point calibration3 points
    Automatic standard solutions
    recognition
    Universal Standard,
    GB Standard
    Cell constant set
    Temperature compensation
    coefficient set
    Salinity compensation coefficient set
    Compensation modeNon-compensatory, linear, pure
    water
    Automatic temperature compensation
    Manual temperature compensation
    Dissolved oxygen MeasurementMeasurement principlePolarographic
    Zero calibration
    Air-Saturation calibration
    Automatic temperature compensation(0.0~45.0) ℃
    Automatic atmospheric pressure
    compensation
    (60.0~110. 0)mbar
    Manual atmospheric pressure
    compensation
    (60.0~110. 0) mbar
    Atmospheric pressure unitsBar, kPa, mbar, Torr, Atm
    Manual salinity compensation(0.0~50.0) g/L
    Temperature
    Measurement
    Temperature units℃, ℉

    Table 7

    1.4. Manual Introduction

    This manual is for the BMET-602 multi-parameter analyzer series, including the use and operation instructions of all measurement functions guidance.

    2. Safety Notices

    Please read the manual before use. The user MUST use the instrument following this manual to avoid damage to the user and equipment.
    Before using the meter, READ the following notes:
    DO NOT DISASSEMBLE the device for inspection or repair. To prevent electric shock or damage to the device, DO NOT place cables and connectors in any liquid, wet or corrosive environment. Please use the defaulted power adapter, DO NOT use it if the power cord is damaged (the wire is exposed or broken). DO NOT use in flammable and explosive environments. DO NOT use if the user finds any abnormalities such as damage or deformation of the device.
    The following identifier will be used in this manual.
    【TIPS】
    Tips help users use the meter.

    3. Terms Explanation

    pH/pX
    pH/pX Slope: The amount of potential change generated by each 1 pH/pX change, expressed in mV/pH or by 100% Theoretical Slope (PTS). pX = - log[X], where [X] means molar concentration (mol/L) of X ions. E0 of pH: Also known as "zero potential", it usually refers to the potential value at a pH of 7. One-point calibration: Calibration with a standard solution. Two-point calibration: Calibration with two standard solutions. Multi-point calibration: Calibration with more than two standard solutions. Unit conversion:
    n pX=-log(10-nmol/L), n>0, 1mol/L=103mmol/L, 1ppt=103ppm=1g/L 1ppm=103ppb=1mg/L, 1ppb=10-3ppm =1μg/L
    Electrical conductivity
    Cell Constant: Also known as the conductivity cell constant. The ratio of the distance to the area of the electrode sheet, is expressed in cm-1. Usually, there are conductance electrodes with several cell constants such as 0.01, 0.1, 1.0, 10, etc. The conductance electrode with a cell constant of 1.0 is the most used one and has a wide measurement range. Temperature Coefficient: The change in conductivity caused by a 1°C change in temperature is usually expressed in %/°C, and the default is 0.02, which is 2.00%/°C. TDS Conversion Factor: The conversion factor between conductivity and TDS, which defaults to 0.71. Unit conversion:1ppt=103ppm=1g/L, 1ppm=103ppb=1mg/L
    1ppb=10-3ppm =1μg/L
    Dissolved oxygen
    Dissolved oxygen concentration: The content of oxygen dissolved in water. Expressed in milligrams of oxygen per liter of water, usually denoted as DO. Dissolved Oxygen Saturation: The ratio of the on-site dissolved oxygen concentration to the saturated dissolved oxygen concentration under the same conditions. Salinity: Salt content in water, expressed in g/L. When the salinity increases by 1g/L at 15℃, the saturated dissolved oxygen of water decreases by about 0.0559 ppm. Zero-point calibration: Electrodes are calibrated in "oxygen-free water" (freshly formulated 5% sodium sulfite solution). Air-Saturation calibration: The electrodes are calibrated in air or water fully dissolved and saturated with air. Atmospheric pressure compensation: The atmospheric pressure affects the measurement of dissolved oxygen concentration and dissolved oxygen saturation, and atmospheric pressure compensation is required. Before calibration, the on-site atmospheric pressure needs to be input, expressed in mbar, and the default is 101.3 mbar.

    4. Overview and Installation

    4.1. Overview


    Figure 1
    Overview-Front View
    1 Meter Body
    2 Display
    3 Power Key
    4 Function selection keys
    5 Electrodes Stand
    6 pH Electrode

    Figure 2
    Overview- Back View
    7 Conductivity socket
    8 DO/T socket
    9 Ground socket
    10 pH electrode socket
    11 Reference electrode socket
    12 RS232 socket
    13 USB socket
    14 Power cable socket

    Figure 3
    Electrodes and connectors
    15 Four pin aviation connectors
    16 pH electrode connectors
    17 Electrode protection cap
    18 pH electrodes
    19 Conductivity electrodes
    20 Conductivity electrode protection cap
    21 Dissolved Oxygen Electrode protection cap
    22 dissolved oxygen electrodes
    【TIPS】
    The BMET-602 series Benchtop Multiparameter Analyzer adopts new electrode connectors.

    Connector Specifications

    Electrode typeConnector
    specifications
    Electrodes Connection
    pH electrodeBNC(Q9)pH electrode, ORP electrode, ISE probe
    DO/T electrodeFour-pin aviationDO electrode, ATC probe
    Conductivity electrodeFive-pin aviationConductivity electrode
    Reference electrodeBananaReference electrode
    Grounding--

    Table 8

    4.2.Instrument Installation

    4.2.1.Electrodes Stand Installation

    Figure 4
    Electrode Stand Installation
    1) Pull out the electrode holder fixing plate on the right side of the instrument,
    2) Insert the multifunctional electrode holder support into the vertical shaft of the multifunctional electrode holder drawer,
    3) Tighten the set screw on the lower part of the pole of the electrode holder.
    4.2.2. Electrodes Connection
    4.2.2.1. Connection of pH electrodes
    Push the pH Electrode into the electrode holder. Remove the socket protector cap of the pH electrode. Connect the pH electrode into the right socket. If the ATC probe is applied, or ATC has been integrated into the pH probe, please connect the ATC probe onto the DO/T electrode socket.
    【TIPS】
    The BMET-602 Series adopted a pH combination Electrode with a four-pin aviation connector. For replacement purchasing, please choose electrodes with the proper connector.

    4.2.2.2. Connection of conductivity electrodes
    Push the conductivity electrode into the electrode holder. Remove the protector cap of the conductivity electrode. Connect the conductivity electrode into the right socket. Combination conductivity probes integrated with ATC probe. If the separate ATC probe is applied, please connect the ATC probe onto the DO/T electrode socket. At the measurement, please choose the right input of ATC in the meter when you applied a separate ATC probe.
    【TIPS】
    The BMET-602 Series adopted a Conductivity Electrode with a five-pin aviation connector.

    4.2.2.3. Connection of dissolved oxygen electrodes
    Push the dissolved oxygen electrode into the electrode holder. Remove the protector cap of the dissolved oxygen electrode socket. Connect the dissolved oxygen electrode into the right socket. All dissolved oxygen electrodes are integrated with the ATC probe.
    【TIPS】
    ➢ The BMET-602 Series adopted DO Electrode with a four-pin aviation connector.
    ➢ The ATC module of the DO electrode as temperature info input also works for pH measurement. As well, the ATC of the connector of pH combination is not available at this period.

    5. Instrument Operation

    5.1 Switch On/Off

    Press and to switch on the meter. The startup screen shows the software version and other related information. After the self-test program, the screen turns to the homepage and the meter is ready to measure.
    The meter is equipped with 8 function keys. Users press and hold the key for more than 3 seconds and release it to shut down.

    5.2 Screen Icons


    Figure 5
    Screen icons explanation
    1 Title. 2 System time. 3 Reading state. 4 Measurement box. 5 Calibration
    information. 6 Measurement box. 7 Measurement box. 8 the user ID. 9 Sample ID.10 Operation explanation.11 Function buttons.
    The instrument displays symbol identification that has the following functional implications:
    Symbol Explanation

    No.SymbolExplanation
    1Reading status, display the measurement status of reading, stable, locked each indicates that the
    processing, stable, and reading completed.
    2The percentage slope of the pH electrode calibration data
    3The Standard buffer solution for calibration
    4The time of calibration
    5The standards are used for calibration
    6Automatic temperature compensation
    7Manual temperature compensation
    8Cell constant
    9The reference temperature in EC measuring
    10Temperature ecoefficiency
    11EC compensation mode
    12Type of cell constant calibration
    13TDS factor
    14Air pressure compensation in DO measuring
    15Salinity compensation in DO measuring
    16User ID
    17Sample ID
    18Operation notice

    Table 9

    5.3 Function Key


    Figure 6
    Function keys explanation
    Key Function Explanation

    No.KeyExplanationNote
    1PowerPress to switch on/off
    2SettingSet the parameters and settings
    3CancelCancel the operation
    4EnterConfirm the option
    5F1Function key, Corresponds to the function options on the screen
    6F2Function key, Corresponds to the function options on the screen
    7F3Function key, Corresponds to the function options on the screen
    8F4Soft function keys, corresponding to the functions on the screen

    Table 10

    5.4 Parameter Settings

    In the measuring, users can set the instrument parameters by pressing "Setting" to set the measuring parameters.
    5.4.1 Tutorial settings
    For the first use, please follow the guide to settings the measurement parameters. After all the settings, press the "Enter" to return to the previous page.
    5.4.2 Select parameters
    The instrument contains up to 3 measurement modules, and each measurement function can select different measurement parameters (Switch the parameters by clicking on the blank area of measurement box on screen). The detail is shown in the following table:
    Instrument Measurement Parameters

    ModulesMeasurement parameters
    pH/pX measurement functionpH, mV, ORP, pX, ion concentration
    Conductivity measurement functionResistivity, conductivity, TDS, salinity, ash
    Dissolved oxygen measurement functionDissolved oxygen current, dissolved oxygen
    concentration, saturation
    Temperature measurement functionTemperature

    Table 11
    Select the proper measurement parameters to perform a measurement. The screen displays the detail of three measurement parameters when three modules are chosen. Two parameters are shown as two modules are chosen. A single parameter is shown by one measurement method being selected.

    Figure 7
    Three-parameter measurement

    Figure 8
    Two-parameter measurement
    5.4.3 Reading Mode Settings

    Figure 9
    Single-parameter measurement
    The meter provides three reading modes, including continuous reading, auto reading, and timed reading.
    Continuous reading: The instrument displays real-time measurement results. Users can end the measurement at any time and save the last result. Auto-reading: The measurement reached the balance, and the meter locked the reading result. The meter offers "Fast", "Medium", "Strict" and "Custom" four options for endpoint detection conditions. Time reading method: Timed Reading contains two kinds of timed reading methods: "Interval Measurement" and "Timed Measurement". "Interval Measurement" provide measurement results at interval time and "Timed Measurement" provide measurement result after a set time.
    Reading Parameters Settings

    Stability TypepHISE/pXConductivityDO
    FastStable time4s4s5s5s
    Fluctuation1mV0.6mV1.0%4nA
    MediumStable time6s8s8s8s
    Fluctuation0.5mV0. 2mV0.4%3nA
    StrictStable time8s12s15s15s
    Fluctuation0.1mV0.1mV0.1%2nA
    Custom
    (Recommended value)
    Stable time1 to 30s1 to 30s1 to 30s1 to 30s
    Fluctuation0.1~1mV0.1~1mV0.1~2%2 to 5nA

    Table 12
    5.4.4 pH Parameter Settings
    5.4.4.1 pH standard groups
    The meter provides various Standards Group including DIN, NIST, GB, and USA. And allows the user to prepare the customized Standard groups.
    Standard Solution Groups

    GroupsContents
    GB1.68pH, 3.56pH, 4.00pH, 6.86pH, 7.41pH, 9.18pH, 12.46pH
    DIN1.68pH, 2.00pH, 3.56pH, 3.78pH, 4.01pH, 6.87pH, 7.00pH, 7.42pH, 9.18pH,
    10.01pH, 12.45pH
    NIST1.68pH, 4.01pH, 6.86pH, 7.00pH, 7.42pH, 10.01pH, 12.47pH
    USA1.680pH, 4.010pH, 7.000pH, 10.010pH

    Table 13

    Figure 10
    Selection of standard groups and standard solution
    The meter supports up to five-points calibrations. Neighboring standards (pH gap<2) choice in the group may be frozen for accurate calibration. For neighboring standards, please choose the customization to perform calibration.
    【TIPS】
    If the selected standard solution group is different from the pH standard buffer solution used, it will lead to wrong calibration results.

    5.4.4.2 Recognition
    Auto Mode and Manual Mode.
    In some special cases, it is necessary to use some non-standard pH buffer solutions, or use two very close pH standard buffer solutions for electrode calibration. In this case, the manual standard solution identification function can be used. When set to "manual mode", the pH value of the current standard solution can be input during and used for electrode calibration.
    5.4.4.3 Resolution settings
    The pH measurement resolution of the instrument is adjustable. pH Resolution: 0.01pH and 0.1pH. mV resolution: 0.1 mV and 1 mV.
    5.4.5 pX/ISE Parameter Settings
    5.4.5.1 Ion mode selection
    The ISE measurement methods provide by meter include Ag+, Na+, K+, NH4+, Cl-, F-, NO3-, BF4-, CN-, Cu2+, Pb2+, Ca2+, etc., also support customized ions.
    5.4.5.2 Resolution settings
    The PX/ISE measurement resolution of the instrument is adjustable. pX resolution: 0.01pX and 0.1pX.mV resolution: 0.1 mV and 1 mV.
    5.4.6 Conductivity Parameter Settings
    5.4.6.1 EC calibration type
    EC calibration type: Cal by Standards and input manually.
    CAl By Standards: Cell constant is calibrated with standard conductivity standard solution.
    Input manually: It allows the user to set the cell constant.
    5.4.6.2 Manual standard recognition
    For neighboring standards, please choose the customization to perform calibration.
    5.4.6.3 Conductivity electrode type
    Conductivity electrode: Four conductivity cell constant 0.01, 0.1, 1, 10. The defaulted conductivity cell constant is 1. Users need to enter the cell constant value on the label of the conductivity electrode for accurate measurement.
    The electrode constant corresponds to the measurement range of conductivity, resistivity and TDS

    No.Cell ConstantRange
    Conductivity Range0.01cm -1(0.000~19.99)μS/cm
    0.1cm -1(0.200~199.9)μS/cm
    1.0cm -1 (platinum black)2.000μS/cm~199.9mS/cm
    10cm -120.00μS/cm~1000mS/cm


    Resistivity Range0.01cm -120.0MΩ·cm~50.0kΩ·cm
    0.1cm -12.00MΩ·cm~5.00kΩ·cm
    1.0cm -1 (platinum black)200.0kΩ·cm~5.00Ω·cm
    TDS Range0.01cm -1(0.000~9.99)ppm
    0.1cm -1(0.100~99.9)ppm
    1.0cm -1 (platinum black)1.000 ppm~99.9 ppt
    10cm -110.00 ppm~300 ppt

    Table 14
    5.4.6.4 EC Reference temperature
    Conductivity reference temperature: The conductivity of the solution is greatly affected by temperature, to make the conductivity measurement results at different temperatures comparable, the conductivity and temperature values at the time of measurement are usually recorded and converted into the conductivity value at a certain temperature through temperature compensation, which is the reference temperature. The instrument allows settings of 20.0°C,25.0 °C 2 reference temperatures, the default reference temperature is 25 °C.
    5.4.6.5 EC Compensation
    EC Compensation Mode: Three different compensation modes can be used for various applications. The meter supports Linear type, DI water type, and non-comp type.
    1) Linear type: Linear compensation is usually used for the measurement of medium and high conductivity solutions. With linear compensation, you can set the temperature compensation coefficient, which defaults to 2.00%/°C (approximately the temperature compensation coefficient of a sodium chloride solution at 25°C). It allows the user to set the temperature coefficient.
    2) DI water type: DI water compensation is usually used for the measurement of pure water and ultrapure hydropower conductivity below 5μS/cm. It allows the user to set the temperature coefficient.
    3) Non-comp type: Non-compensation is usually used to obtain the true conductivity value at the measured temperature.
    5.4.7 Dissolved Oxygen Parameter Settings
    5.4.7.1 DO salinity compensation settings
    Salinity is the amount of sodium chloride dissolved in 1 L of water, in g/L. The dissolved oxygen concentration of water is highly affected by salinity. Typically, for every 1 g/L increase in salinity, the saturated dissolved oxygen of water decreases by 0.0559 mg/L or ppm.
    Users can set salinity compensation on the setting page. The salinity compensation range of measurement is (0.0~50.0) g/L.
    【TIPS】
    If the sample has high salinity, salinity compensation is required to obtain accurate results.

    5.4.7.2 Atmos compensation
    Atmospheric pressure is an important factor for dissolved oxygen concentration and dissolved oxygen saturation measurement. The meter supports automatic barometric compensation and manual barometric compensation modes in the range of (600~1100) mbar. The default atmospheric pressure is 1013.0mbar in manual compensation mode.
    Users can select the necessary compensation mode and pressure unit for measurement in the setting.
    【TIPS】
    Atmos compensation is important for measurement results in low atmospheric pressure areas.
    5.4.8 Temperature Parameter Settings
    The temperature unit of the meter is selectable in °C and °F. Temperature compensation mode: ATC and MTC.
    ATC means automatic compensation. It allows the user to select the 4 pins socket or 5 pins socket.
    MTC means manual compensation. It allows the user to input the temperature.
    【TIPS】
    ATC input selection. For single-parameter measurement, please follow the above instruction. When conductivity compensation electrode dissolved oxygen electrode is applied, the default ATC input is from DO electrode. If the ATC input selection is incorrect, try resetting or manually entering the temperature.
    5.4.9 Data Management Settings
    5.4.9.1 Sample ID type
    The instrument supports three setting methods of Sample ID: number order, time order, and manual.
    Number order: The sample ID No. is increasing with series number. Time order: The sample ID No. is increasing with sample measuring time. Format: Year/Y, Month/M, Day/D, Hour/H, Minutes/M, Second/S Manual: Manually set the sample ID No. It allows samples to manually enter the sample ID when saving or printing data.
    5.4.9.2 Result Autosave
    When this function is enabled, the meter saves the results when the reading is stable in the auto-reading and interval timed reading mode.
    5.4.9.3 Data Overwrite

    The meter provides 500 sets of measurement results storage space. When this function is enabled, the results data that exceeds capacity will overwrite the old results data.
    5.4.10 Output option
    The data format is GLP, STD Format, and Custom. It could select one data format to output the result.
    5.4.11 The user ID Settings
    Set the user ID.
    5.4.12 System Parameter Settings
    5.4.12.1 System Date & Time
    Settings of system date and time.
    5.4.12.2 Buzzer setting
    Users can set the key sound by this setting.
    5.4.12.3 Brightness setting
    Users can adjust the screen brightness by this setting.
    5.4.12.4 Auto Power off
    The meter provides an auto shutdown function. When the meter is not used, the meter switches off automatically.
    5.4.12.5 Restore Default
    The meter supports "Restore Default" and "Restore Parameters". "Restoring Default" will restore all meter parameters to the factory state. "Restoring parameters" will restore the measurement parameters to the factory state.
    5.4.12.6 Software version
    Users can find the software version information on the general setting page.

    5.5 pH Measurement

    5.5.1 Calibration Preparation
    The electrode slope and zero potential of pH electrodes drift slightly over time. To accurately measure pH, it is recommended to calibrate the pH electrode before use, the instrument supports 1-5 points calibration.
    One point calibration is a calibration process with a single standard solution, commonly applied in a quick test. The calibration slope is 100% here.
    Two-point calibration is to use two pH standard buffer solutions to calibrate the electrode and construct a linear calibration curve through two points. Two-point calibration is the most commonly used calibration method, and it is usually recommended that the pH value of the solution be measured lies between the two standard buffer solutions. Two-point calibration can improve pH measurement accuracy.
    Multi-point calibration is a calibration process with more than one standard solution. It is recommended to calibrate between two standard buffer solutions at the pH of the solution to be tested. Multi-point calibration covers a wider measurement range for accurate pH measurement. Before starting calibration, please prepare one or more pH standard buffer solutions.
    5.5.2 Standards group selection
    Before starting calibration, please prepare one or more pH standard buffer solutions. The meter has a standards recognition function. Please set the Standard Group before the measurement.
    You can also set the identification type to "Manual Mode" and manually enter the nominal value during the calibration process.
    5.5.3 pH Calibration
    The calibration process is as follows:
    1. Setting.
    1) Set the parameters (e.g., pH).
    2) Select NIST standard solution group, and check pH 4.01, pH 7.00, and pH 10.01 three standard solutions.
    3) Set the Auto Mode recognition.
    2. Connect the ATC probe or enter the temperature manually.
    3. Press the F2 "Calibrate" for one-parameter measurements or press the F2 "Calibrate"- "pH Calibration" for multi-parameter measurements.
    4. Put the cleaned electrode into a pH 4.01 standard solution.
    5. Wait for the instrument to display “Auto Mode Matched”, and press the F4 “Start”.
    6. If only 1-point calibration is required, after 1-point calibration is completed, press the "Enter" key to complete the calibration.
    7. If multi-point calibration is required, please replace the pH7.01 and pH10.01 standard buffer solutions. After cleaning the electrode, place the electrode into the standard solution. After the instrument recognizes it successfully, the instrument reads stably, press the F4 "Next Point" to complete the calibration.
    8. After completing the calibration, press the "Enter" key to complete the calibration, save the calibration results and end the calibration, and directly enter the start interface. If the checked standard solution group is 5, automatically end the calibration after five points of calibration.
    5.5.4 pH Measurement
    After the calibration, press "Measure" to start a measurement.

    Figure 11
    pH measurement information
    The measurement process is as follows:
    1. Setting.
    1) Set the parameters (e.g., pH).
    2) Set the reading mode (e.g., continuous reading, auto-reading, or timed format).
    2. Put the electrode into the test solution under test.
    3. In the idle status, press F4 "Measure" to enter into measurement status.
    4. When the reading is stable, press "Enter" to read the results.
    5. Press the "Save" to save the measurement results.
    6. Press the "Output" to print the measurement result when connect to the printer.
    7. Between measurements, stored pH electrode in distilled or deionized water.
    8. After measurement, rinse the pH electrode with deionized water thoroughly.
    【TIPS】
    The measurement end of the electrode should well be immersed into the sample solution. For high accuracy measurement, make sure the measurement is carried out at the lab with constant temperature and pressure. If the two temperatures are different, it is recommended to use a pH combination electrode with temperature compensation or use a separate temperature electrode for automatic temperature compensation. Or use a thermometer to measure the temperature of the current solution and manually set the temperature for manual compensation.

    5.6. Ion Measurement

    5.6.1. Calibration Preparation
    The accuracy of the electrode slightly decreases over time. For an accurate measurement, electrode calibration is necessary. See the electrode instructions for specification.
    5.6.1.1 Ion-selective electrodes
    The ion-selective electrode is based on the ion-selective membrane, which can be divided into a single crystal membrane, salt membrane, glass membrane, and PVC ion-selective membrane. Ion-selective electrodes usually have single electrodes and composite electrodes, composite electrodes are more convenient and simpler in the operation. The meter requires a composite electrode.
    5.6.1.2 Ionic strength adjustment buffer
    The use of ion electrodes to measure ion concentration requires the addition of an ionic strength adjustment buffer.
    The ionic strength of a solution has an important influence on the measurement of ion concentration. On the one hand, the ion-selective electrode directly measures the activity of the ion, α = γc. Wherein, α is the activity of the ion, γ is the activity coefficient of the ion, and c is the ion concentration. Usually, the activity coefficient γ is affected by the ionic strength in the solution. By adding an ionic strength adjustment buffer to the standard solution and the test solution, the measured solution has a similar ionic strength to the standard solution, thereby having a similar activity coefficient γ. On the other hand, if a solution with low ionic strength, the potential of the reference electrode will show instability. The addition of the ionic strength adjustment buffer can help stabilize the reference electrode.
    Various ion measurement needs various ionic strength adjustment buffer. Common ionic strength adjustment buffers are recommended in the following table.
    Recommended ionic strength adjustment buffer

    Ions’ categoryionic strength adjustment buffer
    Na +0.2 mol/L diisopropylamine
    F -0.1 mol/L NaCl or TISAB


    Cl -0.1 mol/L KNO 3
    Br -0.1 mol/L KNO 3
    I -0.1 mol/L KNO 3
    Ag +0.1 mol/L NaNO 3
    Cu 2+0.1 mol/L NaNO 3
    Pb 2+0.1 mol/L KNO 3
    S 2-0.1 mol/L KNO 3
    K +0.05 mol/L MgAc 2
    Ca 2+0.1 mol/L KCl
    NO 3 -0.1 mol/L NaH 2 PO 4
    BF 4 -0.1 mol/L Na 2 SO 4
    ClO 4 -0.1 mol/L NaCl

    Table 15
    5.6.1.3 Preparation
    Please prepare standards according to ISO guidelines or buy locally.
    5.6.1.4 Activation of the ISE electrodes
    When the electrode is used for the first time or has not been used for a long time, an activation is recommended. The electrode has better measurement performance after activation.
    Ion electrode activation solution and activation time recommendation

    Ion categoryActivation solutionActivation time
    Na +10 -3 mol/L NaCl2h
    F -10 -3 mol/L NaF2h
    Cl -10 -3 mol/L KCl2h
    Br -10 -3 mol/L NaBr2h
    I -10 -3 mol/L NaI2h
    Ag +10 -3 mol/L AgNO 32h
    Cu 2+10 -3 mol/L Cu (NO 3 ) 22h
    Pb 2+10 -3 mol/L Pb (NO 3 ) 22h
    S 2-10 -3 mol/L AgNO 32h
    K +10 -3 mol/L KCl2h
    Ca 2+10 -3 mol/L CaCl 22h
    NO 3 -10 -3 mol/L NaNO 32h
    BF 4 -10 -3 mol/L NaBF 42h
    ClO 4 -10 -3 mol/L NaClO 42h

    Table 16
    【TIPS】
    The activation time may variously be based on various activation solutions. See the ion-selective electrode manual for specifications.

    5.6.1.5 Stirrer setting
    The flow state of the solution influences the electrode potential of the ion-selective electrode. To improve the stability and repeatability of the measurement, it is recommended to use a stirrer to keep the flow rate of the solution stable during calibration and measurement.
    5.6.2 Select Ion Type
    The meter provides ion measurement methods of Ag+, Na+, K+, NH4+, Cl-, F-, NO3-, BF4-, CN-, Cu2+, Pb2+, Ca2+. Combination ISE electrodes are necessary.
    For an ion method not provided, a new method needs to be created. Press "Create", and fill in the information of ion to finish the ion method customization.

    Figure 12
    Ion mode selection
    5.6.3 Measurement Mode Selection
    Select the ion measurement methods by the user's needs. In general, when the direct-reading method is used for ion concentration measurement and pX measurement.
    Direct Reading Measurement is a commonly used ion concentration measuring method. The Direct Reading mode uses the following Nernst formula to calculate the concentration:
    Ex E0 S log(Cx Cb )
    Wherein,
    Ex~ Equilibrium potential of the sample solution, in mV.
    E0~ zero potential value, in mV.
    S ~ Electrode slope (%).
    Cx~ Concentration value of the sample solution, in mol/L.
    Cb~ Blank concentration value in mol/L.
    The electrode slope and zero potential value can be known from the calibration. When measuring the sample solution, the sample concentration result can be calculated from the formula. The direct-reading method is fast and suitable for quick test measurement of common samples.

    Figure 13
    Direct reading method of pX/ISE measurement
    The calibration setting process is as follows:
    1) Press "pX para setup" to change the "Conc Meas Mode". Select "Direct Reading".
    2) Set "Conc Unit", "Blank Conc", "Conc Resolution" and "mV Resolution".
    5.6.4 pX/ISE Calibration
    The pX/ISE calibration process is as follows:
    1. Add an appropriate amount of standard solution (usually 100 ml) to the beaker, then add an ionic strength adjustment buffer. Adjust the stirring speed of the solution for measurement.
    2.Press the F2 "Calibrate"-"pX Calibration".
    3.Put the cleaned electrode into a standard solution.
    4.Press the F2 “STD value” to input the standard value of the standard solution.
    5.Wait for the reading to be stable, and press the F4 “Start”.
    6.If only 1-point calibration is required, after 1-point calibration is completed, press the "Measure" key to complete the calibration.
    7. If choosing multi-point calibration (up to 5), press "Next Point" to repeat the operation.
    8. If the checked standard solution group is 5, automatically end the calibration after five points of calibration.

    Figure 14
    pX/ISE calibration information
    【TIPS】
    ➢ Please re-calibration for an unexpected measurement result.
    ➢ A room temperature test solution is recommended.
    ➢ It is recommended to calibrate from low concentration to high concentration standards.
    5.6.5 pX/ISE Measurement
    The measurement process is as follows:
    1.Setting
    Set the parameters (e.g. pX) Set the ion type (e.g. F-) Set the direct reading as concentration means mode. Set the con concentration unit (e.g. ppm) Set the reading mode (e.g. continuous reading, auto-readind, or timed format).
    2.Add an appropriate amount of standard solution (usually 100 ml) to the beaker, then add an ionic strength adjustment buffer. Adjust the stirring speed of the solution for measurement.
    3.In the idle status, press F4"Measure" to enter into measurement status.
    4.When the reading is stable, read the results.
    5.Press the "Save" to save the measurement results.
    6.Press the "Output" to print the measurement result when connect to the printer.
    7. Between measurements, stored ISE electrode in distilled or deionized water.
    8. After measurement, rinse the ISE electrode with deionized water thoroughly.

    Figure 15
    PX/ISE measurement information
    After measuring a high ion concentration sample, the low concentration
    ions measurement will be affected. In this situation, it is recommended to rinse the electrode as follows:
    Add an appropriate amount of standard solution (usually 100 ml) to the beaker, then add an ionic strength adjustment buffer. Adjust the stirring speed of the solution for measurement. Rinse the ion-selective electrode with DI water, dry out and place it in a blank solution and place it in a blank solution to wait for the potential to stabilize. After taking out the electrodes, ready to perform the measurement.
    【TIPS】
    Different ISE probes have different potential values in a blank solution. If the blank potential is away from the reference value, the user can do an activation to improve the performance of electrodes. If the electrode still does not meet the requirements, a new electrode is quite considerable.
    【TIPS】
    Please re-calibration for an unexpected measurement result. A room temperature test solution is recommended. It is recommended to calibrate from low concentration to high concentration standards.

    5.7 Conductivity Measurement

    5.7.1 Cell Constant Input
    Conductivity electrodes are precisely calibrated at the time of manufacture and marked with the exact cell constant. Before the measurement, by "Parameter Setting"- "EC Parameter" to enter the electrode cell constant.

    Figure 16
    EC measurement information
    5.7.2 Calibration Preparation
    In general, conductivity electrodes need few calibrations. When the user gets an unexpected result, an electrode calibration is considerable.
    Usually, a single standard solution is required for calibration. For accurate measurement of sample conductivity above 50mS/cm, a two-point calibration is required. Two standards are required, a low conductivity standard and a conductivity standard close to the sample.
    The meter provides various Standards Group including Universal
    Standard Group and GB group. And allows the user to prepare the customized Standard groups.
    5.7.3 Conductivity Calibration
    After the measurement parameters are selected, press "Calibrate".

    Figure 17
    Electrode cell constant calibration information
    For conductivity electrodes with different cell constants, it is recommended to use the following conductivity standard solutions for calibration.
    KCl standards to electrode cell constants

    Cell constant (cm -1 ).0.1110
    KCl solution Concentration (mol/L).0.0010.01 or 0.10.1 or 1

    Table 17
    The calibration process is as follows:
    1.In the idle status, press the soft button F1 "Setting" -"EC Parameter".
    2.Press the "Constant type" to select the "1”.
    3.Press "Cal. Type" to select the "Cal by Standards".
    4.Prepare one or more standard conductivity solutions (e.g.,1413μS/cm conductivity solution).
    5.Prepare a thermostatic bath, and set the temperature to (25.0±0.1) °C.
    6.Place a standard conductivity solution in a thermostatic bath, and set the temperature to (25.0±0.1) °C.
    7.Place the conductivity electrode into a standard solution.
    8. When the conductivity and temperature reading (e.g.,1413µS/cm, 25.0℃) are stable, press the "Start".
    9.If choosing one-point calibration, press "Enter" to end the calibration.
    10.If choosing multi-point calibration (up to 3), press "Next Point" to repeat the operation.
    11.The meter saves calibration data automatically and turns to idle status.
    【TIPS】
    If the standard is not recognized, please check the connection of the probe and the contamination of standards.
    【TIPS】
    The conductivity of the solution is greatly affected by temperature, it is recommended to use constant temperature water for calibration. Automatic or manual temperature compensation can also be optional when there is no water bath.
    5.7.4 Conductivity Measurement

    Figure 18
    Conductivity measurement information
    The measurement process is as follows:
    1.Setting.
    1)Set the parameters (e.g., conductivity). 2) Set the reading mode (e.g., continuous reading, auto-reading, or timed format). 3) Set the EC compensation (e.g., Linear compensation, temperature compensation coefficient 2.00%/°C). 4)Set the reference temperature (e.g., 25°C).
    2.Rinse the conductivity electrode with DI water, and dry out.
    3.Put the measurement end of the electrode into the sample solution.
    4.In the idle status, press the soft button F4 "Measure" to enter into measurement status.
    5.When the reading is stable, read the results.
    6.Press the "Save" to save the measurement results.
    7.Press the "Output" to print the measurement result when connect to the printer.
    8.Between measurements, stored EC electrode in distilled or deionized water.
    9.After measurement, rinse the EC electrode with deionized water thoroughly and put it on the electrode protection cap.

    5.8 TDS Measurements

    5.8.1 TDS conversion factor
    5.8.1.1 Low Concentration TDS Sample
    For samples with relatively simple composition and low concentration, the TDS of the solution can be estimated by conductivity. Compared with the weighing method, TDS estimation by conductivity is relatively simple and convenient with quite a good accuracy. The conversion factor between conductivity and TDS defaulting to 0.71, which can be used as the TDS coefficient for approximate estimation in most situations.
    The conversion factor adjust process is as follows:
    1.Press soft button F1 "Setting" -"TDS Parameter".
    2.Select the TDSF CAL Type as the set TDS Factor.
    3.Input the TDS factor as the desired TDS coefficient.
    Conductivity to TDS Standard Solution

    Conductivity μS/cmTDS standards
    KCl(mg/L)NaCl(mg/L)442(mg/L)
    2311.610.714.74
    8440.3838.0450.5
    447225.6215.5300
    1413744.7702.11000
    1500757.1737.11050
    2070104510411500
    276413821414.82062.7
    8974510144877608
    128807447723011367
    150008759853213455
    80000521684838479688

    Table 18
    442 indicated the solution contains 40%Na2SO4,40%NaHCO3,20%NaCl. The values listed in the table are values at 25°C.
    5.8.1.2 High Concentrations TDS Sample Measurement
    For samples with simple components and higher concentrations, such as high concentrations of NaCl solution, TDS factor re-calibration is needed.
    For TDS measurements, the user may need to correct the TDS conversion factor by the TDS standard.
    The conversion factor calibration process is as follows:
    1.Setting.
    Set the parameters (e.g., TDS). 2) Press soft button F1 "Setting" -"TDS Parameter". 3) Select the TDSF CAL Type as the set Cal by STD. 4) Set the reference temperature (e.g., 25°C).
    2. Prepare TDS Standard.
    3. Place a standard conductivity solution in a thermostatic bath, and set the temperature to (25.0±0.1) °C.
    4.Rinse the conductivity electrode with DI water, and dry out.
    5.Put the measurement end of the electrode into the sample solution.
    6.Press the F2 "Calibrate"-"TDS Calibration".
    7.Set the STD value as the sample STD value.
    8.When the TDS and temperature readings (708ppm, 25.0℃) are stable.
    9.If choosing one-point calibration, press "Enter" to end the calibration.
    10. If choosing multi-point calibration (up to 3), press "Next Point" to repeat the operation.
    11. The meter saves calibration data automatically.

    Figure 19
    TDS coefficients calibration
    5.8.1.3 Complex TDS Sample Measurement
    For samples with complex compositions, the accuracy of TDS measurements can be improved by re-determination by laboratory methods and manual input of TDS coefficients. When the composition or concentration of the sample to be measured changes significantly, it is recommended to recalibrate the TDS coefficient.
    The conversion factor calibration process is as follows:
    1.Rinse the electrode with DI water. Put the measurement end of the electrode into the sample solution and set the temperature at (25.0±0.1) °C.
    2.Using a weighing method to determine the TDS.
    3.Calculate the TDS coefficient.
    4.Press soft button F1 "Setting" -"TDS Parameter".
    5.Select the TDSF CAL Type as the set TDS Factor.
    6.Input the TDS factor as the desired TDS coefficient.
    5.8.2 TDS Measurements
    Users can switch the measurement parameter to TDS measurement by pressing the Conductivity/TDS measurement box on the screen.

    5.9. Salinity Measurement

    The instrument can be used to determine the salinity of sodium chloride. The salinity of sodium chloride can be used to approximate the salinity of the solution being measured. By measuring the conductivity of the sample, the mass percentage of the corresponding sodium chloride solution can be calculated to convert the sodium chloride salinity.
    Users can switch the measurement parameter to salinity measurement by pressing the Conductivity/TDS measurement box on the screen. The detail refers to the measurement method of conductivity salinity measurement.

    5.10 Resistivity Measurement

    Resistivity and conductivity are reciprocal to each other, and conductivity can be measured at the same time when measuring resistivity.
    Users can switch the measurement parameter to resistivity measurement by pressing the Conductivity/TDS measurement box on the screen. The detail refers to the measurement method of conductivity measurement.

    5.11 Dissolved Oxygen Measurement

    5.11.1 Calibration Preparation
    5.11.1.1 Preparation
    Zero oxygen is calibrated using a freshly formulated 5% sodium sulfite solution as oxygen-free water.
    5.11.1.2 Electrodes preparation
    When the DO electrode (polarographic) is used for the first time or has not been used for a long time, the filling solution needs to be replaced.
    Follow the below instruction to replace the filling solution:
    Take the cap off the electrodes, rinse the cap with DI water and dry out. Rinse the inner electrode with DI water and dry the electrode. Add the filling solution (electrolyte) into the membrane cap up to 3/4. Install the cap onto the electrode.
    【TIPS】
    Membrane: The most important part, can't be damaged or lead to incorrect measurement. Electrolyte: Needs to be replaced every 2 weeks to 2 months, depending on the frequency of use.

    5.11.1.3 Electrode Polarization
    Polarographic DO electrodes need to be polarized before every use.
    Connect the DO electrode to the meter and turn on the meter, wait for 1 hour and the electrode are auto polarized. When the electrodes are unplugged from the meter for no more than 1 hour, measurements are allowed after 25 minutes of polarization.
    5.11.2 DO Calibration
    The calibration process is as follows:
    1.Press “Calibrate” to select “DO Calibration”.
    2.Place the electrode into an oxygen-free solution.
    3.When the reading is stable, press “CAL Zero” to do zero oxygen calibration.
    4.Rinse the electrode with DI water again, and place the probe in the upper part of a bottle filled with air-saturated (well shaken) water.
    5.When the reading is stable, press “Cal Air-Sat” to do full oxygen calibration.
    6.After calibration, press “Enter” to end the calibration.
    【TIPS】
    The calibration process ends before it is completed, and the set parameters are not saved.


    Figure 20
    Dissolved oxygen calibration information
    5.11.3 DO Measurement
    The measurement process is as follows:
    1.Setting.
    Set the parameters (e.g., DO). Set the reading mode (e.g., continuous reading, auto-reading, or timed format). Set the salt compensation. Set the Atmos compensation.
    2.Put the electrode into the test solution under test and shake the electrode gently in a circle, in a circular motion, or use a stirrer to avoid air bubbles during the process.
    3. In the idle status, press F4 "Measure" to enter into measurement status.
    4. When the reading is stable, read the results.
    5. Press the "Save" to save the measurement results.
    6. Press the "Output" to print the measurement result when connect to the printer.
    7. Between measurements, stored DO electrode in distilled or deionized water.
    8. After measurement, rinse the DO electrode with deionized water thoroughly. Add the filling solution (electrolyte) into the membrane cap up to 3/4, and Install the cap onto the electrode.

    Figure 21
    Dissolved oxygen measurement information.
    5.11.4 DO Saturation Measurement
    Dissolved oxygen saturation refers to the ratio of the dissolved oxygen concentration to the saturated dissolved oxygen concentration under the same conditions.
    Users can switch the measurement parameter to Saturation measurement by pressing the DO measurement box on the screen.

    5.12 Data Management

    Press "Data" to view the detail of the results.
    The meter stores the measurement results independently according to the measured parameters. The meter provides data Storage 500 sets for each parameter (pH/pX/ISE/EC/Resistivity/TDS/Salinity/DO/Saturation).
    The user can press "Delete" into the delete menu. It allows users to select the current data or all data to delete.
    The user can view the data filter by parameter, and locate No. or stored date. By the filter setting, press "Start Search" to look up the data. The filter data shows in a graph. Press "←" and"→" to choose data. the user can choose one and press the "Enter" key to see the detailed result. Users can press "Delete" to delete the current result. Press "Output" to select the output data. The format supports the output of the current result, output matched result, and output all results.

    Figure 22
    Results setting view

    Figure 23
    Results setting view
    【TIPS】
    Please connect the meter with the printer when the meter is switched off. The Baud rate is 9600bps for printer connecting. The default setting is 8 data bits, one start bit, one stop bit, and no parity.

    The output format is as followed:
    Report Title
    ----------------------------------------
    Measure Time:2021/02/06 13:31:51
    Operator: Admini
    Model: BMET-602 Multi-Parameter
    Analyzer Serial Number:
    SW Version: Ver 1.00
    ----------------------------------------
    .............................SAMPLE INFO
    Sample ID:
    Sample 1
    ..........................ELECTRODE INFO
    pH EC ID: pH Electrode
    ..............................CALIB INFO
    Calib Operator:
    REX Team Calib Time:
    2020/06/18 12:13:10 Calib Num:
    3
    ............................CALIB RESULT
    STD 1: 4.01pH 176.8mV 25.0c
    STD 2: 7.00pH 0.0mV 25.0c
    STD 3: 10.01pH -178.1mV 25.0c
    pH Slope 1:
    100.00% pH E0 1:
    0.0mV pH Slope 2:
    100.00% pH E0 2:
    0.0mV
    ..............................BRIEF INFO
    Reading Mode: Continuous Reading
    Stable Type:
    Medium Temp Comp Type:
    ATC
    ..................................RESULT
    Result:
    7.00pH Signal Value:
    -0.2mV Temp Value:
    25.0c

    6. Maintenance/Troubleshooting

    6.1 Meter Maintenance

    The correct use and maintenance of the instrument can ensure the accurate and reliable performance of the instrument. Additionally, exposure to chemicals or harsh use environments can affect performance.
    The pH/pX electrode socket has a protective plug, when the meter is not in use, please insert the protective plug into the pH/pX socket.
    If the meter is not used for a long time, please disconnect the power supply. The electrode socket of the instrument must be kept clean and dry, and should not be in contact with acid, alkali, and salt solutions. Keep the meter and accessories clean and away from acids, alkalis, and any corrosive solutions/gases. Users can clean the meter surface with clean water and detergent. When the meter is transported, please follow the instructions:
    Please remove all connected cables. Please remove the electrode holder. Please use the original packaging in the long-distance transport to avoid damage.

    6.2 Electrodes Maintenance

    For more detailed information, please refer to the electrode instruction manual.

    6.3 Troubleshooting

    Troubleshooting

    PhenomenonProbable reasonsSolutions
    1. No DisplayDamage to the meter.Replace or repair as required.
    2. Incorrect mV measurement isThe electrode is out of service life The electrode plug is in poor contactReplace the electrodes Connect the protection plug, if the potential is not 0mV, please contact the after-sales service.
    3. Incorrect pH measurementRefer to as 2.2 Refer to as 2.2 The electrodes are not calibrated or are calibrated
    incorrectly
    Refer to as 2.2 Refer to as 2.2 Recalibrate the electrode or replace the standard solution
    4. Incorrect pX/ISE measurement isRefer to as 2.2 Refer to as 2.2 The electrodes are not calibrated or are calibrated incorrectly Incorrect ISE probeRefer to as 2.2 Refer to as 2.2 Recalibrate the electrode or replace the standard solution Buy the correct ISE probe
    5. Incorrect conductivity measurementThe electrode is out of service life The electrodes are not calibrated or are calibrated
    incorrectly
    Replace the electrodes Recalibrate the electrode or replace the standard solution
    6. Incorrect DO measurementThe electrode is out of service life The electrodes are calibrated incorrectly Stirring speed is not goodReplace the electrodes Recalibrate the electrode Keep the bubbles in the stirring state consistent

    Table 20
    If the meter still does not work, please contact your local dealer for further assistance.

    7. Technical Supports

    Accessories

    Please refer to the accessories table for purchasing recommendations.
    Meter accessories

    NameDescription
    E-301-QC 3 in 1pH composite electrodepH Measurement Probe
    DJS-1VTC conductivity electrodeConductivity, TDS Measurement Probe
    DO-958-Q dissolved oxygen electrodeDissolved Oxygen Measurement Probe
    Standard pH buffer solution 4.01/7.00/10.01Standard solution
    Conductivity solution 1413μs/cmStandard solution

    Table 21

    8. Appendixes

    Appendix 1

    pH-Temperature Relationship Table of pH Standard Solutions

    Temperature(°C)1.684.017.0010.01
    51.674.007.0910.25
    101.674.007.0610.18
    151.674.007.0410.12
    201.684.007.0210.06
    251.684.017.0010.01
    301.684.016.999.97
    351.694.026.989.93
    401.694.036.979.89
    451.74.046.979.86
    501.714.066.979.83

    Table 22

    Appendix 2

    Conductivity Standard Solution
    Table 1 KCl standards to electrode cell constants

    Cell constant (cm -1 ).0.1110
    KCl solution Concentration (mol/L).0.0010.01 or 0.10.1 or 1

    Table 23
    Table 2 Approximate concentrations of KCl solutions and their conductivity values (μS/cm) relationship.

    T(°C)84μS/cm1413μS/cm12.88mS/cm
    553.028968.22
    1060.3410209.33
    1567.61114710.48
    2075.80127811.67
    2584.00141312.88
    3092,19155214.12
    35100.92169615.39

    Table 24