Home
Biological Safety Cabinet
Class III Biosafety Cabinet
Biological Safety Cabinet Class III BCBS-503
Biological Safety Cabinet Class III BCBS-503
- Sea, Air, Door to Door Shipping
- 1 Year Warranty
- US & European Standards
Biolab Biological Safety Cabinet Class III Series is designed to work with level 4 pathogens. These safety cabinets feature a physical barrier between the user and product. It involves a gas tight chamber with negative pressure and attached pair of gloves for operation to avoid direct contact with hazardous contaminants. Robust design with advanced protection, it has a user friendly interface and energy efficiency.
- With memory function in case of power-failure
- UV lamp ensures work area sterilization effectively
- An angled cabinet front ensure an ergonomic working posture
- Exhaust air is double-filtered through high-quality ULPA filters with typical efficiency of 99.9995% for 0.12μmparticles
Specification
Features
Type | Class III Biosafety Cabinet |
ULPA Filter | Air supply filter: Efficiency 99.9995% at 0.12μm First exhaust filter: Efficiency 99.9995% at 0.12μm Second exhaust filter: Efficiency 99.9995% at 0.12μm |
Pre-filter | Polyester fiber, washable |
Pressure | ≥ negative120 Pa |
Work Zone Material | 304 stainless steel |
Front Window | 8 mm toughened glass, anti-ultraviolet radiation |
Gloves | One pair of 800mm butyl rubber gloves |
UV Lamp | 18W x2 ,8W x1 Emission of 253.7 nanometers for most efficient decontamination |
LED Lamp | 12W x2 |
Illumination | ≥ 1000 lux |
Display | LCD display |
Noise | ≤67dB |
Visual and Audio Alarm | Audio and visual alarm, abnormal pressure; filter replacement |
Standard Accessories | Gloves, fluorescent lamp, UV lamp x3, remote control, pressure meter, drain valve |
External Size (WxDxH) | 1540x880x2000 mm |
Internal Size (WxDxH) | 910x650x660 mm |
Package Size (WxDxH) | 1635x1065x2160 mm |
Pass Box Size(WxDxH) | Internal Size:430x330x355 mm External Size:575x425x495 mm |
Gross Weight | 350 kg |
Consumption | 1200 W |
Power Supply | AC220V±10% ,50/60 Hz;110V±10%,60Hz |
- With memory function in case of power-failure
- UV lamp ensures work area sterilization effectively
- An angled cabinet front ensure an ergonomic working posture
- Exhaust air is double-filtered through high-quality ULPA filters with typical efficiency of 99.9995% for 0.12μmparticles