Water Bath BBWA-105

Thermostatic Water Bath
  • Cold-rolling steel spraying exterior.
  • Stainless steel Inner tank and cover.
  • Intelligent Temperature control instrument adopts PID control, dual digital display, with functions of timing of operation (0-9999 min), fixed value operation, automatically stop, realization of over-temperature alarm, power-off compensation, power failure memory, error correction and Menu lock.
Old SKU: BER1DL5
Specifications Features Manual Get Quote

Specifications

ModelBBWA-105
Voltage200-240V/100-120V
Power(W)1000
Temmotion (°C)±0.5
Tem-range (°C)Room Temp.±5~100°C
Tem. Sensitivity (°C)≤±1
Display error (°C)≤±2.5
Chamber Size (mm)325*325*120
Products Size (mm)510*380*190
Packing Size (mm)550*475*235
N.W.(KG)6.5
G.W.(KG)7.7

Features

Cold-rolling steel spraying exterior.
Stainless steel Inner tank and cover.
Intelligent Temperature control instrument adopts PID control, dual digital display, with functions of timing of operation (0-9999 min), fixed value operation, automatically stop, realization of over-temperature alarm, power-off compensation, power failure memory, error correction and Menu lock.

Operating Manuals

Download Manual (PDF)

1. Safety warning before use
2. Main features of product
3. Main technical parameters
4. Instrument operation and use instructions
5. General fault and troubleshooting methods

1. Safety warning before use

Tags:, Description:, Attributes-End, Mandatory prohibition, •Do not store volatile, flammable or explosive materials in the Mandatory prohibition

•Do not store volatile, flammable or explosive materials in the equipment, otherwise it may cause explosion or fire.

•Do not place the equipment in a place that is wet, wet or may splash water, otherwise it may lead to accidents such as leakage, short circuit or electric shock.

•Non-professional technicians shall not disassemble, repair or modify the equipment, otherwise it may cause fire or electric shock accidents due to improper operation.


Obligatory item, The following items may cause injury, equipment damage, and related property losses:, • The equipment Obligatory item

The following items may cause injury, equipment damage, and related property losses:

• The equipment should be installed on a solid ground. If the ground is not solid enough or the installation site is not suitable, injuries may be caused by the equipment tipping over.

• Please use the special power supply marked on the nameplate of the equipment. The power socket must be connected with ground. The equipment must be reliably grounded to prevent accidental leakage and electric shock or fire.

• Before any repair or maintenance of the equipment, be sure to disconnect the power supply to prevent electric shock or personal injury.

• Be sure to wear gloves when performing equipment repair or maintenance to prevent injury caused by touching edges or sharp corners.

• If the equipment is found abnormal operation, immediately unplug the power plug, stop the operation of the equipment. Operation under abnormal conditions may cause electric shock or fire.

2. Main features of product

1. the box is made of high quality cold rolled steel plate refining, the surface electrostatic spraying, novel shape, beautiful.

2. the inner liner and top cover are made of high-quality stainless steel, corrosion resistance and aging resistance, long service life.

3. immersion U-type heating tube direct heating, small heat loss, fast heating speed.

4. the use of special instrument temperature control, high precision temperature control.

3. Main technical parameters

Model

BBWA-105

voltage

120-240V/50-60Hz

Temp. Range

Room temperature +5°C~100°C

Temp.fluctuation

±0.5°C

Specification

2 rows

4 holes

Power (w)

1000

Inner Chamber Size (mm)

325x325

x120

Table 1

4. Instrument operation and use instructions

1. The instrument should be placed on a horizontal table surface.

2.add water to the shelf 50mm or more, (if conditions are best to use distilled water) shall not be empty burning or lack of water.

3. Turn on the power switch, the instrument is powered on, the upper row shows InP, the lower row shows the index number, indicating the input type. After 2 seconds, the upper row shows the upper range, and the lower row shows the lower range, indicating the measurement range. After another 2 seconds, the measured value is displayed in the upper row, and the set value is displayed in the lower row, and it enters the normal working state.

Obligatory item, The following items may cause injury, equipment damage, and related property losses:, • The equipment

Figure 1

Temperature setting: Press the function key, the upper row displays SP, press ▲ press ▼ key, so that the lower row displays the required set temperature value. Press the Set key to display the ST constant temperature time setting, and then press the Set key to return to the standard mode.

If appears on the lower side of the top row of digital tube display, it means that the thermocouple is reversed, and HHH appears on the upper side, it means that the thermocouple/thermal resistance is open or the temperature exceeds the measurement range.

Boiling point function: When selecting the water bath, set the temperature >100 degrees, heat the full output and make the water boil.

Each function parameter table:

PROMPT

Name

Set range

Notes

Initial value

AL

Upper limit alarm

0... Measuring range °C

Alarm setting, alarm insensitive zone is 0.4 fixed value

10.0

AL2

Lower limit alarm

0... Measuring range °C

Alarm setting, alarm insensitive zone is 0.4 fixed value

-5.0

Pb

Measurement correction

Full scale range

The measurement error caused by the sensor and thermocouple compensation wire is corrected

00.0

P

Belt of proportion

1.0…300°C

Proportional action regulation, the larger P is, the smaller proportional action is, and the lower system gain is

2.8

I

Integration time

20…999seconds

The integral action time constant, the larger I, the weaker the integral action

90

d

Differential time

0---999seconds

The differential action time constant, the larger D, the stronger the differential action, and can overcome the overshoot, D=0PI control

20

T

Cycle of control

1…100seconds

The relay outputs 20s,SSR and SCR switch 3s

10

PK

Self tuning

0,1

0 off self tuning 1 on self tuning

0

LK

Combination lock

0,1,2

0: no lock, 1: lock parameters except the set value, 2: lock all parameters

0

Table 2

5. General fault and troubleshooting methods

phenomenon

Causes and elimination methods

No power supply

① Turn on the switch, turn off the light is not bright, the instrument does not show that the plug and socket contact is bad or the switch is bad should check the plug, socket or switch.

②The switch light is on, the instrument has no display, the instrument is bad to replace the instrument.

Not heating up

① Set the temperature low and water temperature, adjust the set temperature.

② Instrument or sensor is bad, change the instrument or sensor.

③ The heating tube is bad, change the heating tube.

The temperature keeps rising out of control

Instrument out of control or sensor short circuit, change the meter or sensor.

Display temperature

It's too different from the actual temperature

Fix the SC zero position parameter

The temperature keeps rising out of control

Instrument out of control or sensor short circuit, change the meter or sensor.

Display temperature

It's too different from the actual temperature

Fix the SC zero position parameter

Table 3