Key Features
- These purifiers are designed to provide water for clinical analyzers specified to Clinical Laboratory Reagent Water standard (formally Clinical Laboratory Standards Institute Type). Our product range delivers consistent water quality feeding a single clinical analyzer, large automated analyzers and multiple analyzers on a distribution loop.
- Re-circulation always ensures desirable pure water quality. UV, micro-filtration technology, and easy sanitization ensure optimal bacterial purity of less than 1 CFU/ml.
- An emergency bypass prevents downtime by providing an uninterrupted supply of pure water to the analyzer at all times. We understand that downtime cannot be allowed in the clinical laboratory.
SKU: BFN-11801
Pricing & Resources
- Up to 3-Year Product Warranty
- US & EU Compliance Standards
- Door-to-Door Shipping
Recently Shipped To:
🇲🇲 Myanmar , 🇱🇰 Sri Lanka , 🇳🇱 Netherlands , 🇹🇼 Taiwan
Get Quote
Download Catalog
Technical Specifications
Detailed operating and construction parameters for this model.
| Model | BFN-11801 |
| Production Rate | 30L/hr |
| Water Quality | Type II |
| Feed Water | Tap Water |
| Tank Capacity | |
| Feed Water Requirement : | |
| Source | Tap water |
| Conductivity | <2000 μS/cm |
| Hardness | <450 ppm as CaCO₃ |
| Pressure | 0.1-0.5 MPa |
| Temperature | 5-40°C |
| High Quality Purification Water (Class II) : | |
| Resistivity at 25°C | ≥10 MΩ·cm |
| TOC | <30 ppb |
| Dissolved Organic | <0.1 ppm |
| Particulate (≥0.02 μm) | <1 pc/ml |
| Bacteria | <1 CFU/ml |
| Flow Rate | 1.0-1.5 L/min |
| Electrical Requirements : | |
| Electrical Voltage | 110V/220V ±10% |
| Electrical Frequency | 50Hz/60Hz |
| Packing Information : | |
| Net Weight Main Units | 25 kg |
| Net Weight Water Tank | 5 kg |
| External Dimensions (W x D x H) Main Units | 315 x 525 x 570 mm |
| External Dimensions (W x D x H) Water Tank | 380 x 380 x 595 mm |
| Shipping Weight Main Units | 37 kg |
| Shipping Weight Water Tank | 13 kg |
| Shipping Dimensions (W x D x H) Main Units | 525 x 610 x 770 mm |
| Shipping Dimensions (W x D x H) Water Tank | 520 x 440 x 615 mm |
| Notes | * If feed water quality is poor (Conductivity >1000 μS/cm), a 3-class strengthened pretreatment module and RO-2 type are highly recommended.** When the hardness of feed water is high (>450 ppm as CaCO₃), a 0.5T water softener tank is recommended. |
Product Features
Key design, performance, usability, and safety advantages.
These purifiers are designed to provide water for clinical analyzers specified to Clinical Laboratory Reagent Water standard (formally Clinical Laboratory Standards Institute Type). Our product range delivers consistent water quality feeding a single clinical analyzer, large automated analyzers and multiple analyzers on a distribution loop.
Re-circulation always ensures desirable pure water quality. UV, micro-filtration technology, and easy sanitization ensure optimal bacterial purity of less than 1 CFU/ml.
An emergency bypass prevents downtime by providing an uninterrupted supply of pure water to the analyzer at all times. We understand that downtime cannot be allowed in the clinical laboratory.