Get Price
Single Beam Scanning UV Visible Spectrophotometer BFP-501
Uv/vis Spectrophotometer
Key Features
- Scan type, single beam spectrophotometer with wide wavelength range, satisfying requirement of various fields.
- Five options for spectral bandwidth selection: 5nm, 4nm, 2nm, 1nm, and 0.5nm, made according to customer's need and satisfy the requirements of pharmacopoeia.
- Standard manual 4-cell holder accommodates cells from 5-50mm and changeable to long path length cell holder of 100mm.
- Optional accessories such as peristaltic pump automatic sampler, water constant temperature sample holder, peltier temperature control sample holder, single slot test tube sample holder, film sample holder.
- Optimized optics and electronics design, light source and detector from the world famous manufacturer ensure high performance and reliability.
- Rich measurement methods: wavelength scan, time scan, multi-wavelength determination, multi-order derivative determination, double-wavelength method and triple-wavelength method etc., meet different measurement requirements.
- Data output can be obtained via a printer port.
- Parameters and data can be saved in case of power failure for user's convenience.
SKU: BFP-501
Pricing & Resources
- Up to 3-Year Product Warranty
- US & EU Compliance Standards
- Door-to-Door Shipping
Technical Specifications
Detailed operating and construction parameters for this model.
| Model | BFP-501 |
| Wavelength Range | 190-1100nm |
| Spectral Bandwidth | 2nm(5 nm, 4 nm, 1 nm, 0.5 nm optional) |
| Wavelength Accuracy | ±0.5nm |
| Wavelength Reproducibility | ≤0.2 nm |
| Monochromator | Single beam, plane grating of 1200 L/mm |
| Photometric Accuracy | ±0.3%T (0-100%T) |
| Photometric Reproducibility | ≤0.2%T |
| Working Mode | T, A (-0.301-3A), C, E |
| Stray Light | ≤0.1%T (NaI, 220 nm; NaNO₂, 340 nm) |
| Baseline Flatness | ±0.003A |
| Stability | ≤0.002A/h (at 500 nm, after warming up) |
| Noise | ≤0.2%T/3 min (0% line) |
| Detector | Silicon photodiode |
| Display | 7 inches colorful touch screen |
| Power | AC: 90-250V, 50V/60Hz |
| Dimensions | 470 x 325 x 220 mm |
| Weight | 9 kg |
Product Features
Key design, performance, usability, and safety advantages.
Scan type, single beam spectrophotometer with wide wavelength range, satisfying requirement of various fields.
Five options for spectral bandwidth selection: 5nm, 4nm, 2nm, 1nm, and 0.5nm, made according to customer's need and satisfy the requirements of pharmacopoeia.
Standard manual 4-cell holder accommodates cells from 5-50mm and changeable to long path length cell holder of 100mm.
Optional accessories such as peristaltic pump automatic sampler, water constant temperature sample holder, peltier temperature control sample holder, single slot test tube sample holder, film sample holder.
Optimized optics and electronics design, light source and detector from the world famous manufacturer ensure high performance and reliability.
Rich measurement methods: wavelength scan, time scan, multi-wavelength determination, multi-order derivative determination, double-wavelength method and triple-wavelength method etc., meet different measurement requirements.
Data output can be obtained via a printer port.
Parameters and data can be saved in case of power failure for user's convenience.
PC controlled measurement can be achieved with a USB port for more accurate and flexible requirements.