Oil Bath Circulator BCOB-101
- Microcomputer intelligent control system: high temperature circulator, rapid heating, stable temperature, easy to operate.
- Water and oil dual use: the highest temperature can reach 300°C.
- LED double Windows display the temperature measurement value and temperature setting value respectively, and the touch button is easy to operate.
- Large flow rate of the external circulation pump, up to 20 L/min, 12 L/min and 8 L/min.
- The cold water circulation device can be selected to realize the rapid cooling of the system through the cold source access, which is suitable for the temperature control of the heat release reaction at high temperature.
- Suitable for chemical reaction requiring high temperature temperature field, biological medicine and other industries.
- Optional magnetic mixing devices can be used for high-temperature reactions.
- A vertical constant temperature oil tank is equipped with universal casters, which can be easily handled, suitable for jacket reaction kettle, chemical pilot reaction, high temperature distillation, semiconductor industry.
Specifications
| Model | BCOB-101 |
| Studio Size (mm) | 250x200x200 |
| volume | 10L |
| Temperature control range | Room temperature of ~300°C |
| temperature fluctuation | ±0.1°C |
| groove depth (mm) | 200 |
| pumping flow (L/min) | 8 |
| Voltage (V) | 220 |
| rough weight (KG) | 12.3 |
| net weight (KG) | 11.13 |
Features
Water and oil dual use: the highest temperature can reach 300°C.
LED double Windows display the temperature measurement value and temperature setting value respectively, and the touch button is easy to operate.
Large flow rate of the external circulation pump, up to 20 L/min, 12 L/min and 8 L/min.
The cold water circulation device can be selected to realize the rapid cooling of the system through the cold source access, which is suitable for the temperature control of the heat release reaction at high temperature.
Suitable for chemical reaction requiring high temperature temperature field, biological medicine and other industries.
Optional magnetic mixing devices can be used for high-temperature reactions.
A vertical constant temperature oil tank is equipped with universal casters, which can be easily handled, suitable for jacket reaction kettle, chemical pilot reaction, high temperature distillation, semiconductor industry.
Vertical structure, the temperature control part can be removed separately, can be carried for other constant temperature environments as a heating source, but also easy to replace the temperature control accessories.
Applications
Operating Manuals
1. Introduction
1.1 Application
1.2 Features
1.3 Performance Characteristics
1.4 Technical parameter
2. Control Panel
3. Operate
3.1 The order of call-out of each function
3.2 Detailed description of each function
3.3 Self-toting function of the controller
3.4 Temperature control parameter setting
1. Introduction
1.1 Application
The oil bath is suitable for direct and auxiliary heating in scientific applications requiring precise temperature control, such as biology, chemistry, physics, botany, and chemical engineering.
1.2 Features
This oil tank uses oil as the heating medium. It is specifically designed for the inspection and comparison of glass thermometers, and is suitable for use in laboratories, temperature measurement departments, and thermometer manufacturing factories.
1.3 Performance Characteristics
The temperature control is achieved through digital display instruments, making the operation very convenient:
The temperature adjustment is carried out using a continuous PI.D method, ensuring precise temperature control. The inner container is made of stainless steel, which is durable. The outer shell is made of thin steel plate with plastic coating, and it adopts a forced circulation method of the medium to ensure uniform temperature distribution.
The stainless steel electric heating tubes are selected, which are easy to maintain.
1.4 Technical parameter
Heating power: 1kW
Power supply voltage: 220V 50Hz
Temperature range: RT + 5 - 300°C
Temperature accuracy: ±0.1°C
2. Control Panel

Figure 1
1. Wind Fast: Show the percentage of fan output;
2. Appointment time indicator light: it will be on when the appointment is turned on;
3 .Time: fixedValue timingTime;
4 .Key lock;
5 .Heating power;
6 .Measure the temperature;
7 .Set the temperature;
8 .Alarm indicator light: flash when it is down or abnormal;
9 .The operation is bright ;
10. Stop Bright: It is on during down time;
11 .water indicator light (standby);
12. lighting indicator light (standby);
13. heating indicator light;
14. refrigeration indicator light (standby);
15. water level indicator light flashes when there is a shortage of water (standby);
16.Fan indicator light: turn on when the fan is output;
°C Symbol: When the temperature unit is Celsius°CBright,
℉ Symbol: Fahrenheit
℉ Bright;
Button Explain
:Set Metal bolt;
:Shift key;
: Reduce the key; : 
3. Operate
3.1 The order of call-out of each function
Power-on work of the controller:
After the controller is powered on, the monitor Fully lights up, and then the upper limit of the range and the lower limit of the range are displayed. After another 2 seconds, the measurement value is displayed in the upper row, the set value is displayed in the lower row, and enters the standard display mode.
Temperature,Time, wind speed Settings:
Press the SET key, SV flashes, set the temperature, press or key, make the display as the desired temperature value. Press the function key again, the time flashes (unit is minute), set the timing time (if the timing time is not required, the value is 0), press or to display the required time value, and press the function key again, show F(fan), press OR key, so that the display is the required value, then press the function key, the controller returns to standard mode
3.2 Detailed description of each function
PV temperature measurement value; SV Temperature setting value
If "
" is displayed, the sensor is open or the input signal exceeds the measurement upper limit. If"
"is displayed, the sensor is short-circuited or the input signal is lower than the lower limit of the measuring range. When the input signal of the controller is not within the measuring range, the buzzer buzzes. Press any key to mute the buzzer.
Timer function:
When time is set to 0, the timing function of the controller is cancelled and the controller keeps running. When time is not set to 0, the controller provides the timing function. When the running time of the controller arrives, the lower row displays END, the buzzer chirps, the controller stops working, and press any key to silence, and at the same time Push downKey andKey 4 seconds,it can be restarted.
3.3 Self-toting function of the controller
Long press the key for 4 seconds and then release, the instrument starts the self-setting function, and the running light flashes. When the temperature runs back and forth for several times, the lamp becomes steady on, and the instrument self-setting ends (set the setting time to 0, cancel the timing, and set the timing time again after the end of the self-setting).
3.4 Temperature control parameter setting
In the standard state, long press the SET key, the LCD screen will display LK, adjust LK=18,click the SET key to enter the temperature setting.
5 groups of PID:
Related parameters: segmentation point F1~F4, PID1~PID5
Set the value <= F1, execute PID1
Set value [F1~F2], execute PID2
Set value [Fahrenheit~F3], execute PID3
Set value [Fahrenheit3~Fahrenheit4], execute PID4
Set value > F4, execute PID5
Multiple corrections:
Related parameters PK1 to PK8 are modified in more points.
Modified method, the measured value of the temperature point is directly input to the corresponding PK position.
If a pair of corrected points is no longer needed, the PK is adjusted to the minimum value, which is -1999.
The table below for each functional parameter.
Sign of Prompt | Name | Setting Range | Illustration | Initial Value |
Pb | Zero adjustment / Intercept | -100~ 999 -100.0~ 999.9 | When the zero error of the controller is large and the error of full scale is small , adjust this value. This value is rarely adjusted for PT100. | 0 |
P K | Full scale adjustment / Slope | -1000~ 2000 | When the zero error of the controller is small and the error of full scale is large, adjust this value. PK = 4000 x (reading value of mercury thermometer − current temperature value) / current temperature value. In general, adjust this value first for PT100. | 0 |
rH | Upper range setting | 0~400°C / 0.0~ 400.0 | Adjust rH to set the range of 0~rH °C. | User redefined |
AL | Alarm setting | 0~full scale / 0.0~full Scale | When the temperature exceeds SP + A L value, the alarm light and alarm will turn on. | 6 |
P | Proportional band | 0~full scale / 0.0~full scale | The larger the P value, the smaller the proportional effect. P = 0 is positional control, and the insensitive area is 0.4. | 15 |
I | Integral time / Reset time | 10~3600 sec | The larger the integral action time constant, the weaker the integral action. | 300 |
d | Derivative time / Rate time | 0~3600 sec | The larger the time constant, the stronger the differentiation. | 100 |
Ar | Overshoot inhibition | 0~200% | The smaller the XF value, the stronger the controller's ability to suppress temperature overshoot. | 100 |
Table 1