Forced Convection Oven BODR-304

Forced Air Drying Oven
  • 1. Horizontal air circularly by double air ducts, high efficiency and high uniformity.
  • 2. PID controller, with over-temperature protection function.
  • 3. With fan switch,LED digital display,high temperature blast motor.
  • 4. New anti-hot handle.
Old SKU: BER1B12
Specifications Features Applications Manual Get Quote

Specifications

ModelBODR-304
Cycle ModeForced convection
Function
--Temp. RangeRT+10-300°C
--Temp. Resolution Ratio0.1°C
--Temp. Motion±1°C
--Temp. Uniformity±2.5%
Structure
--Inner ChamberMirror Stainless Steel
--Quter ShellCold rolling steel electrostatic spraying exterior
--Insulation LayerHigh quality rock wool board(with CE)
--HeaterStainless steel heater
--Power Rating2.3kW
--Exhaust HoleΦ35mm top(with function of test hole)
Controller
--Temp. Control ModeTwo temperature section PID intelligent
--Temp. Setting ModeTouch button setting
--Temp. Display ModeMeasuring temperature display on upper LCD screen, setting temperature display on the lower row
--Timer0-9999min(with timing wait function)
--Operation FunctionOperation Function Fixed temperature operation, timing function, auto stop.
--Additional FunctionHigh precision digital independent temperature limiter.4 grades 3 speed adjustable-speed fan. Sensor deviation correction, temperature overshoot self-tuning, internal parameter locking, power-off parameter memory
--SensorPT100
--Safety DeviceDigital Independent Temp. limiter, Over temperature sound-light alarm.
Specification
--Inner Chamber size(W*L*H)(mm)500*450*550
--Exterior Size (W*L*H)(mm)640*640*930
--Packing Size (W*L*H)(mm)710*730*1100
--Volume125L
--Max. Shelf Number11
--Load Per rack15kg
--Shelf Space40mm
--Power Rating (50/60HZ)AC220V/10.5A
--NW/GW (kg)60/66
Accessory
--Shelf2
--Shelf Frame4
Optional accessoriesSpare shelf, RS485 interface, printer, Program controller, U-disk storage;3.5-inch color touch screen temperature controller
Fan Switches

Features

1. Horizontal air circularly by double air ducts, high efficiency and high uniformity.
2. PID controller, with over-temperature protection function.
3. With fan switch,LED digital display,high temperature blast motor.
4. New anti-hot handle.

Applications

Used for drying, baking, wax, heat treatment.

Operating Manuals

Download Manual (PDF)

1. Safety warning before use
2. Main features of product
3. technical specification
4.Instrument operation and display instructions
2.Operation and usage methods
3.Reference and setting of internal temperature parameters
5. Instrument wiring diagram
6. General faults and troubleshooting methods

1. Safety warning before use

Tags:, Description:, Attributes-End, Must prohibit items, The following matters may cause serious injury or death to Must prohibit items

The following matters may cause serious injury or death to personnel in accidents


1. Do not store volatile, flammable, or explosive materials in this device, as it may cause

explosions or fires.

2.Do not place this device in a place that is exposed to rain, moisture, or may splash water, otherwise it may cause accidents such as leakage, short circuit, or electric shock.

3.Non professional technicians are not allowed to disassemble, repair, or modify equipment, otherwise improper operation may cause fire or electric shock accidents to personnel.


Must comply with items, The following matters may cause personal injury, equipment damage, and related property losses., Must comply with items

The following matters may cause personal injury, equipment damage, and related property losses.

1.This device should be installed on a solid ground. If the ground is not solid enough or the installation location is not suitable, personnel may be injured due to equipment overturning.

2. Please use the dedicated power supply indicated on the nameplate of this device, and the power socket must be grounded. This device must be reliably grounded to prevent accidental electrical leakage that may cause electric shock or fire.

3. Before carrying out any repairs or maintenance on this equipment, be sure to disconnect the power supply to prevent electric shock or personal injury.

4. When repairing or maintaining equipment, it is necessary to wear gloves to prevent personal injury caused by touching edges or sharp corners.

5.If any abnormalities are found in the operation of the device, immediately unplug the power plug and stop the device from running. Operating under abnormal conditions may cause electric shock or fire.

2. Main features of product

1. Unique vertical dual duct circulating horizontal air supply, high drying efficiency, and uniform temperature distribution in the studio.

2. Standard fan switch, high brightness LED instrument operation indicator light, and high-temperature resistant blower motor.

3. Adopting a new type of anti-scald handle, the studio adopts high-quality mirror stainless steel processing to form, and the four corner arc transition is easy to clean.

4. The LCD screen of the temperature controller displays and drives the solid-state relay output to control heating, making it safer and more reliable.

5. A 4-speed 3-speed speed adjustable fan can adjust the required air volume according to different experimental requirements of users, thereby drying various types of samples.

6. Independent over temperature protection system: When the temperature in the studio exceeds the set alarm value of the instrument, the bright LED red alarm light will light up and cut off the main heating circuit, improving the safety of product use.

7. Independent temperature limiter: Digital electronic temperature limiter, set to limit the temperature of the studio, providing dual safety protection for the product.

3. technical specification

Model

BOFC-101

BOFC-102

BOFC-103

BER-104

BOFC-104

BOFC-105

BOFC-106

BODR-301

BODR-302

BODR-303

BER-111

BODR-304

BODR-305

BODR-306

Cycle Mode

Forced convection

Function

Temp. Range

RT+10-300°C

Temp. Resolution Ratio

0.1°C

Temp. Motion

±1°C

Temp. Uniformity

±2.5%

Structure

Inner Chamber

Mirror Stainless Steel

Outer Shell

Cold rolling steel electrostatic spraying exterior

Insulation layer

High quality rock wool board(with CE)

Heater

Stainless steel heater

Power rating

0.8kW

1.2kW

1.6kW

2.0kW

2.3kW

3.0kW

5.0kW

Exhaust hole

Φ35mmtop(with function of test hole)

Rear exhaust

Top double exhaust

Controller

Temp. control mode

Two temperature section PID intelligent

Temp. setting mode

Touch button setting

Temp. display mode

B model: Measuring temperature: display on the four digital tubes upper row, setting temperature: display on the lower row

BE model: Measuring temperature display on upper LCD screen, setting temperature display on the lower row

Timer

0-9999min(with timing wait function)

Operation function

Fixed temperature operation, timing function, auto stop.

Additional function

B model: Sensor deviation correction, temperature overshoot self-tuning, internal parameter locking, power-off parameter memory

BE model: High precision digital independent temperature limiter, 4

grade 3speed adjustable-speed fan (230 and 625model BE without fan speed adjustor),sensor deviation correction, temperature overshoot self-

tuning, internal parameter locking, power-off parameter memory

Sensor

PT100

Specification

Inner Chamber size(W*L*H)(mm)

310*310*310

350*350*350

400*360*450

450*420*450

500*450*550

600*500*750

660*760*1250

Exterior size (W*L*H)(mm)

450*500*690

490*540*730

540*550*830

590*610*830

640*640*930

735*835*1115

902*942*1677

Packing size (W*L*H)(mm)

540*575

*855

580*615

*895

635*625

*995

680*685

*1000

710*730

*1100

857*885*1287

1061*1000*1925

Volume

30L

45L

65L

85L

125L

230L

625L

Shelf number

5

6

8

8

11

17

11

Load per rack

15kg

Shelf space

40mm

100mm

Power rating (50/60HZ)

AC220V/3.6A

AC220V/5.5A

AC220V/7.2A

AC220V/9.0A

AC220V/10.5A

AC220V/13.6A

AC380V/13.2A

NW/GW (kg)

33/37

37/43

44/49

50/56

60/66

94/120

150/190

Accessory

Shelf

2

Shelf frame

4

optional accessories

B model: Spare Shelf;

BE model: Spare Shelf, RS485 Interface, Printer, Program controller, U-disk storage; 3.5-inch color touch screen temperature controller

Table 1

4.Instrument operation and display instructions

1.Instrument operation and usage


Figure 1 BODR-301 BODR-302 BODR-303 BODR-304 BODR-305 BODR-306

Figure 1


Figure 2 BODR-301 BODR-302 BODR-303 BODR-304 BODR-305 BODR-306

Figure 2


Key Definition

1. [SET]: Set button. When the main interface displays the status, click this button to enter

the temperature and time target value setting state. Long press this button for 3 seconds to enter the internal parameter setting state.

2. Key Definition, 1. [SET]: Set button. When the main interface displays the status, click this button to enter, the ]: Shift/self tuning key. Clicking this key in the set state can cause the set value to

shift and flash for modification. In the main interface display state, long press this button for 6 seconds to enter the temperature self tuning selection state.

3. Key Definition, 1. [SET]: Set button. When the main interface displays the status, click this button to enter, the ]: The decrease/rerun button can be used to decrease the set value by clicking or long pressing it in the set state. Press and hold this button for 3 seconds in the main interface and at the end of the run state to restart the operation.

4. Key Definition, 1. [SET]: Set button. When the main interface displays the status, click this button to enter, the ]: Add/lock screen button, click or long press this button in the set state to gradually increase the set value.If the lock screen function is selected, in the main interface display state, click this button to lock or unlock the screen.


Definition of LCD instrument indicator lights:

1. [Main interface] Indicator light: This light will light up in normal working state (not set

state), and turn off in reverse.

2. [Lock] indicator light: It lights up in the lock screen state, and goes off in the opposite

direction.

3. [AT] indicator light: During the temperature self tuning process, this light flashes,

otherwise it goes out.

4. [Alarm] Indicator light: When there is a temperature deviation alarm or abnormal

temperature measurement, this light will turn on. When there is a temperature deviation

alarm, this light will flash. Under normal conditions, this light will turn off.

5. [Heating] Indicator light: When there is heating output, this light will turn on, and vice

versa, it will turn off.

6. [A] Indicator light: This light flashes during the reservation timing process, and goes off if

not.

7. [RUN/STOP] indicator light: STOP will only light up after the timer ends, and RUN will

light up in other states.

8. [↑/→/↓] Indicator lights: They flash respectively during heating, constant temperature,

and cooling.


Definition of digital tube display type instrument indicator lights:

1. [AT/F] indicator light: This light comes on when the temperature unit is Fahrenheit, flashes during the temperature self tuning process, and goes off in other states.

2. [TIM] indicator light: It lights up when there is a timing setting, flashes during the reservation timing process, and goes off when there is no timing function.

3. [OUT] indicator light: When there is heating output, this light will turn on, and vice versa, it will turn off.

4. [LOC] indicator light: lights up in lock screen mode, otherwise goes off.

5. [SIN] indicator light: Invalid reservation.

6. [ALM] indicator light: When there is a temperature deviation alarm or abnormal temperature

measurement, this light will turn on. When there is a temperature deviation alarm, this light will flash. Under normal conditions, this light will turn off.

2.Operation and usage methods

In the main interface display state, click the [Set] button to enter the temperature setting state. The PV area displays the prompt SP, and the SV area displays the temperature setting value. The desired setting value can be modified by using the [Shift], [Increase], and [Decrease] keys; Click the 【 Settings 】 button again to enter the time setting state. The PV area displays the prompt ST, and the TIME area displays the time setting value. Use the【 Shift 】 , 【 Increase 】 , and 【 Reduce 】 keys to modify to the desired setting value; Click the [Settings]

button again to exit this setting state, and the set values will be automatically saved.

Explanation: LCD type, "symbol" represents "time unit"; Digital type, with the symbol "TIME display area decimal point".

3.Reference and setting of internal temperature parameters

In the main interface display state, long press and hold the [Set] button for 3 seconds. The password prompt Lc will be displayed in the PV area, and the password value will be displayed in the SV area. Use [Increase], [Reduce], and [Shift] to modify the desired password value. Then click the [Set] button. If the password value is incorrect, the instrument will automatically return to the main interface display state. If the password value is correct, enter the internal parameter setting state, and click the [Set] button to modify each parameter in sequence. During this process, long press the [Set] button for 3 seconds to exit this state, and the parameter values will be automatically saved. Please refer to the table below for details:

Explanation: In the parameter table, the temperature setting value is abbreviated as SP, and the

temperature measurement value is abbreviated as P V.


Parameter Table

Indicator

Parameter Name

Indicator

(Scope) Factory value

Lc

password

When Lc=3, parameter values can be viewed and modified.

0

ALH

upper deviation over temperature alarm.

When PV>SP+ALH, there is an upper deviation over temperature alarm.

(0~100.0°C)

20.0

ALL

Lower deviation over temperature alarm

When PV < SP-ALL, there is a lower deviation over temperature alarm.

Explanation: When ALL=0, the lower deviation alarm is invalid.

(0~100.0°C)

0

Pb

Correction of measurement temperature deviation

Used to correct errors generated during temperature measurement.

Pb=actual temperature value - PV

(-50.0~50.0°C)

0

PL

Correction of

Measurement temperature slope deviation

Usually used to correct errors generated during high-temperature measurements.

PL=1000 * (actual temperature value - PV) ÷ PV

(-999~999) 0

ndT

Timing method

0: No timing; 1: Constant temperature timing; 2: Run timing

(0~2) 1

Tdn

Timing direction

0: Positive timing; 1: Countdown

(0~1) 0

Hn

Timing unit

0: minutes; 1: Hour

(0~1) 0

SPd

Constant temperature deviation

When SP SPd ≤ PV ≤ SP+SPd, enter a constant temperature state

(0.1~50.0°C)

0.5

EST

End Timing

Prompt Time

When the timing ends, the buzzer indicates the time. Note: when the EST = 9999, said a permanent reminder

(0~9999)

60

EH

End Timing

Constant Temperature

Controller

0: Turn off the heating output after timing;

1: Keep constant temperature controlling after timing

(0~1) 0

LF

Lock Screen Function

0: Lockless screen function;

1: Lock screen function, unlock without password.

2: Lock screen function, need password to unlocked.

(0~2) 0

LdT

Lock Screen Delay

In the main screen state, if no key is pressed in the delay LDT time, the controller will automatically lock the screen.

Description: LDT = 600, the delay screen locking function is invalid

(10~600s) 30

PAd

Unlock Password

The password must be entered to unlock it.

(0~9999) 1

Add

Mail Address

Local Address

Description:

PC-E9000 has no communication function.

(1~32) 1

Table 2


Parameter Table

Indicator

Parameter Name

Indicator

(Scope) Factory value

Lc

password

When Lc=567, parameter values can be viewed and modified.

0

rST

factory reset

0: Cancel; 1: Confirm.

(0~1) 0

Table 3

5. Instrument wiring diagram

Table 3, Figure 3 BODR-301 BODR-302 BODR-303 BODR-304 BODR-305 BODR-306

Figure 3

6. General faults and troubleshooting methods

Fault phenomenon

Fault inference

Fault handling methods

No display on startup

The measured temperature is higher than the set temperature or the instrument enters a high temperature alarm state

After turning on, the temperature does not display and does not rise

Fault phenomenon

Power supply not connected

Check for voltage in the power socket

The power plug is not properly plugged in

Check if the contact between the power plug and the socket is reliable

The power switch is not turned on

Turn on the power switch on the right side of the instrument

The fuse on the box is damaged

Replace the power fuse with the same specification

No display on startup

The measured temperature is higher than the set temperature or the instrument enters a high temperature alarm state

The box door is not tightly closed

Close the box door tightly

The instrument has not yet entered a constant temperature state

Wait for a moment before making an observation

After turning on, the temperature does not display and does not rise

The power cord is not properly plugged in

Insert the power cord properly

Sensor malfunction

Notify the factory for repair

The heater is broken

Notify the factory for repair

Table 4