Get Price

Class II B2 Biological Safety Cabinet BKF-801

Bsc-b2 Bio-safety Cabinet
Key Features
  • Air curtain isolation design prevents internal and external cross contamination, with 100% air flow discharged, negative pressure vertical flow, and no need to install pipelines.
  • The front glass can be moved up and down allows for arbitrary positioning which is easy to operate, and complete closure for sterilization. The positioning height limit alarm
  • prompts.
  • The power output socket in the work area is equipped with waterproof sockets and sewage outlets, providing great convenience for operators.
  • HEPA filter is installed at the exhaust area to control emission pollution.
  • The working area is made of high-quality 304 stainless steel, which is smooth, seamless, and has no dead corners. It can be easily and thoroughly disinfected to prevent corrosion and disinfectant erosion.
  • Controlled by an LCD panel with built-in UV light protection device, it can only be opened when the safety door is closed.
  • Equipped with DOP test port and built-in differential pressure gauge.
SKU: BKF-801
Pricing & Resources
  • Up to 3-Year Product Warranty
  • US & EU Compliance Standards
  • Door-to-Door Shipping
Recently Shipped To: 🇳🇬 Nigeria , 🇲🇦 Morocco , 🇲🇿 Mozambique , 🇮🇸 Iceland
Get Quote Download Catalog
Specifications Features Get Quote

Technical Specifications

Detailed operating and construction parameters for this model.

ModelBKF-801
Internal Dimensions (W*D*H)1040*650*620
External Dimensions (W*D*H)1200*800*2270
CleanlinessHEPA: ISO 5 (Class100)
Number of colonies≤0.5pcs/dish·hour (Φ90mm culture plate)
Wind speedAverage suction wind speed: ≥0.55±0.025m/sAverage descending wind speed: ≥0.3±0.025m/s
Filtration EfficiencyHEPA of borosilicate glass fiber material: ≥99.995%, @0.3μm
Noise≤65dB(A)
Vibration half peak≤5μm
PowerAC single phase 220V/50Hz
Maximum power consumption1400W
Weight210KG
HEPA filter specifications and quantity980x490x50x520x380x70
LED/UV lamp specifications and quantity12Wx/20W

Product Features

Key design, performance, usability, and safety advantages.

Air curtain isolation design prevents internal and external cross contamination, with 100% air flow discharged, negative pressure vertical flow, and no need to install pipelines.
The front glass can be moved up and down allows for arbitrary positioning which is easy to operate, and complete closure for sterilization. The positioning height limit alarm
prompts.
The power output socket in the work area is equipped with waterproof sockets and sewage outlets, providing great convenience for operators.
HEPA filter is installed at the exhaust area to control emission pollution.
The working area is made of high-quality 304 stainless steel, which is smooth, seamless, and has no dead corners. It can be easily and thoroughly disinfected to prevent corrosion and disinfectant erosion.
Controlled by an LCD panel with built-in UV light protection device, it can only be opened when the safety door is closed.
Equipped with DOP test port and built-in differential pressure gauge.
10 ° inclination angle, in line with human body design concepts.

Get Quote