Blood Coagulation Analyzer BEM-42401
Blood Coagulation Analyzer
Key Features
- 1. Dual-channel independent detection, can analyze different items at the same time.
- 2. 5-inch high-definition touch screen, easy to operate.
- 3. Open system, good compatibility, Supports reagents of various specifications.
- 4. 20,000 results storage.
- 5. Equipped with multiple pre-temperature positions to improve detection efficiency.
- 6. Built-in thermal printer.
- 7. The pipette automatically triggers the measurement, and the measurement time can be accurate to tenths of a second.
- 8. Less reagent consumption.
SKU: BEM-42401
Pricing & Resources
- Up to 3-Year Product Warranty
- US & EU Compliance Standards
- Door-to-Door Shipping
Recently Shipped To:
🇺🇸 United States , 🇳🇬 Nigeria , 🇱🇧 Lebanon , 🇲🇹 Malta
Get Quote
Download Catalog
Technical Specifications
Detailed operating and construction parameters for this model.
| Model | BEM-42401 |
| Test Principle | Optical colorimetry |
| Test Items | PT, APTT, TT, FIB |
| Test Channel | 2(can test different projects at the same time) |
| Wavelength | 470nm |
| Detection Time | Normal sample detection time is 20 to 40 seconds, and can be set arbitrarily from 20 to 99 seconds |
| Sample Incubation Position | 12 |
| Reagent Incubation Position | 5 |
| Detection Position Temperature | 37.0±1.0°C |
| Incubation Position Temperature | 37.0±1.0°C |
| Incubation Time | 1~999s |
| Reaction Time | 1~99s |
| Reagent Dosage | Minimum dosage 20μl |
| Sample Volume | 20μl~40μl |
| Display | 7-inch high-definition color touch screen |
| Storage | 200,000 results storage, automatically save measurement data when power off |
| Communication Interface | RS232 serial interface, support LIS system |
| Printer | Built-in thermal printer |
| Power Supply | AC100~240V, 50/60Hz |
| Consumption | 70VA |
| Ambient Temperature | 10°C~30°C |
| Relative Humidity | ≤80% |
| Atmospheric Pressure | 86kPa~106kPa |
| External Size(W*D*H) | 360*210*120mm |
| Net Weight | 3.5kg |
| Package Size(W*D*H) | 376*246*316mm |
| Gross Weight | 5kg |
Product Features
Key design, performance, usability, and safety advantages.
1. Dual-channel independent detection, can analyze different items at the same time.
2. 5-inch high-definition touch screen, easy to operate.
3. Open system, good compatibility, Supports reagents of various specifications.
4. 20,000 results storage.
5. Equipped with multiple pre-temperature positions to improve detection efficiency.
6. Built-in thermal printer.
7. The pipette automatically triggers the measurement, and the measurement time can be accurate to tenths of a second.
8. Less reagent consumption.