Get Price
Basic Water Purification System BBPS-302
This basic series is ideal for wide range of applications. It produces RO, Deionized water and Ultrapure water. The organic rejection rate is greater than 99% using reverse osmosis. The resistivity reaches up to 18.2M?.cm which completely meets the highest grade I standard.
- Built-in 20 liters airtight plastic pressure water tank
- Built-in 13 liters high-capacity polishing resin cartridge
- Unique design and easy-to-replace cartridges pack unit.
- Data storage and RS 232/USB communication port.
- Up to 3-Year Product Warranty
- US & EU Compliance Standards
- Door-to-Door Shipping
Technical Specifications
Detailed operating and construction parameters for this model.
| Model | BBPS-302 |
| Water Inlet | Tap water: TDS<200 ppm (Extra pretreatment filter is recommended, if TDS>200 ppm) |
| Temperature | 5-45°C |
| Pressure | 1.0-4.0 Kgf/cm² |
| Flow Procedure | PF+AC+RO+AC |
| Ion rejection rate | 96%-99% (New RO membrane) |
| Organic rejection rate | >99% (when MW>200 Dalton) |
| Particles and bacteria rejection rate | >99% |
| Bacteria | <0.1 cfu/ml (with optional 0.45+0.1μm PES terminal filter) |
| Particles(>0.2μm) | <1/ml (with optional 0.45+0.1μm PES terminal filter) |
| Output (25°C) | 30 L/hrs |
| Pure water outlet | RO water |
| Water Quality Monitor | Portable TDS/conductivity test pen |
| Dimension LxWxH | 410x220x420 mm |
| Weight | 20 kg |
| Standard configuration | Main body (Including 1 set of cartridges) + 15 liters tank + TDS pen + accessory bag |
| Power Consumption (W) | 72 W |
| Power Supply | AC110-220 V, 50/60 Hz |
| Note | *The feed water quality will influence the pure water's quality and cartridges life-span. PF: polypropylene spun fiber, AC: active carbon, RO: reverse osmosis. All specifications are tested under feed water TDS=200 ppm, 25°C, 50 psi and 15% recovery rate. |
| Ultrapure Water Quality | |
| Flow rate | 2.0 L/min (with pressure tank) |
Product Features
Key design, performance, usability, and safety advantages.
Applications Explore common laboratory, research, industrial, and process applications.
Food Industry, Beverage Industry, Agriculture, Boiler Feed, Disaster Relief Environmental, Hospital, Hotel, Marine, Military, Mining, Pharmaceutical, Power
Operating Manual View instructions, operating procedures, safety information, and maintenance guidance.
1. Preface
2. Specification
3. Water Flow Chart
4. Working Environments
5. Installation
5.1 Preparation for installation
5.2 Tube and adapter's connection
5.3 Installation steps
6. Usage Guide
6.1 Starting Up
6.2 Getting Corresponding Pure Water
6.3 Standby
6.4 Shutdown
6.5 The Usage to Keep High Quality Pure Water
7. Water Quality Test
8. Consumables
9. Normal Trouble Diagnosis
10. Warranty & Repair Regulation
1. Preface
Dear customer, in the beginning, we sincerely thank you for choosing our water purification system. This water purification system has incorporated new cutting-edge technology. It is installed and used easily, and can provide you with RO water and ultrapure water for science research. So, it will benefit your work.
For the water purification system's maximum efficiency, it is suggested that the user manual should be read before installation. Any questions in the installation process, please contact our technology engineers or dealers.
2. Specification
Model | BBPS-301 | BBPS-302 | |
Output(25°C)* | 15 liters/hour | 30 liters/hour | |
Flow rate | Up to 2 liters/minute (with pressure tank) | ||
Pure water outlet | 1: reverse osmosis water | ||
RO water quality | |||
Ion rejection rate | 97%-99% (new RO membrane) | ||
Organic rejection rate | >99%, when MW>200 Dalton | ||
Particles and bacteria rejection rate | >99% | ||
Bacteria | <0.1cfu/ml (with optional 0.45+0.1μm PES terminal filter) | ||
Particle(>0.1μm) | <1/ml (with optional 0.45+0.1μm PES terminal filter) | ||
Feed water requirements | Tap water, temperature:5-35°C,pressure:1.0-4.0bar | ||
Dimension and weight | LengthxWidthxHeight:410x220x420mm / Weight: about 16Kg | ||
Electrical requirements | AC110-240V, 50/60Hz | ||
Power | 72W | ||
Standard configuration | Main body (Including 1 set of cartridge)+15 liters pressure tank+ TDS/conductivity test pen | ||
Purification System | |||
Sequence number | Specification | Quantity/set | |
LV.1 | Special spun fiber cartridge | 1 | |
LV.2 | Special active carbon block cartridge | 1 | |
LV.3 | Special active carbon block cartridge | 1 | |
LV.4 | 100GPD RO membrane(15) | 1 | |
200GPD RO membrane(30) | 1 | ||
Table 1
3. Water Flow Chart

Figure 1
4. Working Environments
• Inlet water: Tap water (TDS<200ppm will be suggested).
If inlet water TDS>200ppm, pretreatment is recommended. Water with higher TDS will affect the quality of outlet water and life of purification cartridge.
• Work temperature: 5-35°C
• Pressure: 1.0-4.0bar
• Power: AC100-240,50/60Hz,72W
【Clean, dry working environments would be suggested!】
5. Installation
5.1 Preparation for installation
• The purification system should be installed horizontally and near to tap.
5.2 Tube and adapter's connection
The adapter of the machine is high quality easy-put adapter. And the material of tube is high quality PE.
Tube installation and drag diagram

Figure 2
ATTENTION:
• The tube should be cut with special tube cutter for rounded cut section. And rounded cut section should be guaranteed as much as possible with other cut tools.
• Connect the tube-press the oval cap of the interface strongly, then insert the tube to the bottom of adapter.
• Take off the tube-press the oval cap of the interface strongly, then drag out the tube. Do not drag when the tube can't be dragged out any more.
• The fore-end of the tube, which has been inserted to adapter, should be cut, when it will be used again.
• Sufficient PTFE thread seal tape should be used in all the threaded joints for water leakage inhibitor or preventing.
5.3 Installation steps
(1 ) Open the packing-case, take out main body, accessory box, water tank (optional).
(2) Take out adapters and tube from accessory box, and read the Instruction Manual carefully.
(3) External interface are on the back of machine, and it is labeled with different color's label. Moreover, its adapters are inserted with different color's stop plug.
ATTENTION:
• Stop plug should be pulled out before the following steps.
(4). Connect To Tap Water
There are two ways to connect to the tap.
Guide Chart of 2 Ways to Connect To The Tap

Figure 3
is the place where the interface of machine's inlet water should be connected.
• 1st way-with tap water adapter 1(1/2″internal thread to 1/4″fast-plug) to connect to tap water.
1st. Step: connect tap water adapter 1 to water source
Close the valve of the gooseneck. Dismantle the faucet of gooseneck. Screw tap water adapter 1 into the external thread of gooseneck.
2nd. Step: connect tap water adapter 1 to interface of machine's inlet water
Use 1/4" PE tube with a suitable length. Insert one side into the interface of tap water adapter 1, and insert the other into the interface with blue label marked "To inlet water" at the back of machine.
• 2nd way-with tap water adapter 2(tee joint and 1/4″ball valve) to connect to tap water.
1st. Step: connect tap water adapter 2 to water source
Close the chief valve of tap water. Dismantle the tap.
Screw the 1/4'' ball valve with external thread into the side thread with internal thread of tee joint.
Screw the tap into the internal thread at one end of the tee joint, and at last, screw the other end with external thread of the tee joint(with 1/4'' ball valve and the tap at this time) into the internal thread of the tube, where the tap has been connected.
ATTENTION:
• Sufficient PTFE thread seal tape should be used in all the threaded joints for water leakage inhibitor or preventing
2nd. Step: connect tap water adapter 2(1/4″ball valve) to interface of machine's inlet water
Use 1/4" PE tube with a suitable length. Insert one side into the interface of 1/4″ball valve, and insert the other into the interface with blue label marked "To inlet water" at the back of machine.
ATTENTION:
• Extra pretreatment filters (optional) should be connected between the water source and main body.
(5). Connect To RO Wastewater
Use 1/4" PE tube with a suitable length. Insert one side into the interface with black label marked "To drain" at the back of machine, and the other side is directed to drain. (DO NOT JAM!!)
(6). Connect To Water Tank:
Screw the plastic ball valve onto interface with external thread, which is on the top of pressure tank.
Use 1/4" PE tube with a suitable length. Insert one side into the interface with yellow label marked "To water tank" at the back of machine, and insert the other into the plastic ball valve.
ATTENTION:
• The ball valve on the top of the tank should be open all the time except for maintenance.
Thus the installation is OK.
Installation Guide Chart

Figure 4
6. Usage Guide
All data have been set in the factory.
The machine will operate smoothly without any data-setting and debugging.
The power switch and the switch of pure water outlet is on the side of the shell. Specific picture is as follows :

Figure 5
6.1 Starting Up
Turn on the tap water valve insert the power line into the power source and turn on the power switch then the system begins to produce pure water.
6.2 Getting Corresponding Pure Water
Press the switch to "I", which is on the side of the box, to get RO water. When getting water process is ok, press the switch to "O" position.
6.3 Standby
When RO water and ultrapure water is not for use, the system will be in standby state. The system still produce RO water to store in the water tank(optional). Until tank is full, the system will automatically stop. The system will begin to produce pure water again when any pure water is used.
6.4 Shutdown
Turn off the tap water valve and turn off the power switch. Then it is ok.
ATTENTION:
• Make sure that the source water and power source is not connected, when the system is not in the use state for long time (for example, off duty).
6.5 The Usage to Keep High Quality Pure Water
(1). The pure water is easily polluted by surrounding environment. So getting fresh pure water is suggested.
(2). Keep water tank from sunlight for microbe's reproducing.
(3). When get high pure water, initial high pure water is suggested to drain to get steady pure water.
(4). Avoid air bubble when get pure water to reduce air pollution.
ATTENTION:
• The microbe's reproducing will reduce the life of cartridge, when the machine does not work for long time. So the machine's work every 7-10days is necessary.
7. Water Quality Test
The system has 1 method of water quality measuring- PORTABLE TDS PEN.
• The other is TDS/Conductivity pen. which is used to testing reverse osmosis water and tap water's quality.
Measure unit: TDS(total dissolved solid, ppm) or conductivity(μs/cm)
REMARKS:
• Under normal conditions, new RO membrane's desalination rate is above 95%. It means that TDS of RO water should be less than TDS of inlet tap waterx5%.
• If TDS of RO water > TDS of inlet tap waterx10%, it means that RO membrane's desalination can't meet the minimum requirements. RO membrane should be replaced at once.
• Conversion relations between TDS and conductivity rate(μs/cm):
If TDS<50ppm, conductivity rate (μs/cm) ≈TDSx2
If TDS>200ppm, conductivity rate (μs/cm) ≈TDSx(1.5~1.7).
8. Consumables
Item No. | Specification | Quantity/set | Suggested replacement term |
PC-PP-ZK | Special spun fiber cartridge | 1 | About 2-6 months |
PC-AC-B-ZK | Special active carbon block cartridge | 2 | About 4-6 months |
RO-100GPD | 100GPD RO membrane | 1(15) | About 12-24 months |
RO-200GPD | 200GPD RO membrane | 1(30) | About 12-24 months |
Table 2
REMARKS:
• Worse inlet feed water quality or big dosage will reduce cartridge life.
Consumables guide chart

Figure 5
9. Normal Trouble Diagnosis
Normal trouble | Cause | Diagnosis |
No power | -No plug in -Power adapter broken | -Check the power connecting -Replace new adapter |
No pure water goes out or a little amount of pure water | -Valve of pure water outlet broken -Pump broken -Cartridges or filters' life ends | -Replace new valve -Replace new pump -Replace new cartridges or filters |
Cartridges' life warns | -Cartridges' life ends | -Replace new cartridges |
Water leakage | -Adapter or something broken | -Check, insert and drag out again, replace |
Water quality deteriorate | -Cartridges or filters' life ends -Water quality sensor broken | -Replace new cartridges or filters -Replace new water quality sensor |
Table 3
10. Warranty & Repair Regulation
The products enjoy repair service from the day of purchase. In one year from the purchasing day, we are obliged to replace components for customers free of charge, due to non-human-behavior factors, except for:
(1). All the consumables;
(2). Damage caused by maloperation or use in abnormal situations;
(3). Disassembly any part of the machine or human-behavior damage;
(4). Not repaired by our serviceman.
Specification can be changed without any prior notice for development.