Laboratory water purification system uses double stage reverse osmosis technology. It produces double stage RO water, Deionised, EDI and ultrapure water. These systems have 3 way on-line water quality sensor, multiple alarm with unique design and it has easy-to-replace cartridges pack unit.
  • Integrating with Ionpure Electro deionization technology and module.
  • The largest capacity is 240 liters pure water per day.
  • Automatic microcomputer controlling system, multi-menu operating, real-time animation mode display.
  • Super-large LCD (Resolution:240×128, dimension:106×57mm) display, display the system running state and various parameters intuitively.
Instant Quote
Specification Features Applications
Feed Water Requirements*: 
      Water InletTap water: TDS<200 ppm (Extra pretreatment filter is recommended if TDS>200 ppm)
      Temperature5-45°C
      Pressure1.0-4.0 Kgf/cm²
Flow Procedure**-
Ion rejection rate96%-99% (New RO membrane)
Organic rejection rate>99%,when MW>200 Dalton
Particles and bacteria rejection rate>99%
Output(25°C)****10 L/hrs
Pure water outlet2: RO water, Electro Deionization water
DimensionLxWxH500x360x540 mm
Weight25 kg
Standard configurationMain body (Including 1 set of cartridges) + 20 liters tank+accessory bag
Power Consumption (W)120 W
Power SupplyAC110-220 V, 50/60 Hz
Note*The feed water quality will influence the pure water's quality and cartridges life-span. **PF:polypropylene spun fiber, KDF:kinetic degradation fluxion, AC:active carbon, RO:reverse osmosis, SF:softener, EDI: electro deionization, UV:ultraviolet, TF:terminal microfiltration. ***Value of number will be influenced by temperature and feed water quality. ****All the specifications are tested under the situation:feed water's TDS=200ppm, 25°C, 50psi and 15% recovery rate.
Ultrapure Water Quality: 
      TOC***<30 ppb
Flow procedure**PF+KDF+AC+RO+SF+EDI
EDI water quality: 
      Resistivity***>5 MΩ.cm
      Silicon rejection rate>99.9%
  • Integrating with Ionpure Electro deionization technology and module.
  • The largest capacity is 240 liters pure water per day.
  • Automatic microcomputer controlling system, multi-menu operating, real-time animation mode display.
  • Super-large LCD (Resolution:240×128, dimension:106×57mm) display, display the system running state and various parameters intuitively.
  • 3 way on-line sensor, detect the quality of feed water, RO water, deionized water or ultrapure water respectively.
  • System sterilization procedure, achieve the disinfection of ultrapure water’s pipeline.
  • System circulation function, circulate water when the system stops working, to keep water quality.
  • Self-flushing of the reverse osmosis membrane, extend the life of RO membrane.
  • Multiple alarm functions: such as no water, full water, disqualification of feed water, RO water, deionized water or ultrapure water, cartridges’ life-span ends.
  • The cartridges’ life-span can be set, the time used and left can be displayed, replacing auto-reminding, avoiding the decline of water quality.
  • Level II password, protect all the parameters setting, and prohibit any unauthorized settings change.
  • Water dispensing function-timing and quality (Time range:1-99min, water quality range:0.1-18.2MΩ.cm).
  • RS 232/USB communication port (optional), at least store 1 year's water quality data.
  • Different external tanks (optional) to meet every need and assure ample water-supply.
  • Human engineering design, molding process, high-strength, streamline plastic shell.
  • Pretreatment cartridges, RO module, Electro deionization module, ultrapure cartridges, all designed to modularization independently. Easy
  • to maintenance and replacement.
  • Pipeline and fast-plug adaptor with NSF authorization, assure high quality ultrapure water.
  • KDF pretreating cartridge, replace the ordinary active carbon, prolong the life-span to 12 months, reduce the running cost.
  • DOW’s RO membrane, ensure stable operation and high desalinization rate.
  • 4 in 1 ultrapure cartridges (also can be divided to 4 independent cartridge), with DOW’s nuclear-grade polishing resin, ensure ultrapure water’s quality up to 18.2 MΩ.cm, with the lowest TOC dissolution.
  • Double wavelength (185&254nm) ultraviolet lamp module, restrain bacteria’s increase and reduce TOC.
  • MWCO 5000D ultrafiltration module, effectively eliminate endotoxin precise cell cultivating and IVF.
  • (0.45+0.1)μm double layer PES terminal disinfection filter, assure the quality absolutely axenic.